![Audio Precision DCX-127 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/audio-precision/dcx-127/dcx-127_user-manual_3008577034.webp)
Specifications
DCX-127 Multifunction Module User’s Guide
Page 29
DC VOLTS MEASUREMENT RELATED
Accuracy
1
6 rdg/sec
25 rdg/sec
200 mV range
0.05% + 0.03 mV
0.05% + 0.1 mV
2 V range
0.05% + 0.1 mV
0.05% + 1 mV
20 V range
0.05% + 1 mV
0.05% + 10 mV
200 V range
0.05% + 10 mV
0.05% + 100 mV
500 V range
0.05% + 100 mV
0.05% + 1 V
Resolution
200 mV--200 V ranges
0.005% of range
0.025% of range
500 V range
100 mV
500 mV
Input Resistance
10 MW , ± 1% (all ranges)
Common Mode
Rejection
>120 dB, at dc and 50 Hz to 20 kHz
Common Mode Range
500 Vpeak
Normal Mode Rejection >60 dB, 50 Hz to 60 HZ
1
Valid from + 15°C to +30°C, <80% RH, for 1 year. Derate linearly to
2 times indicated values at + 5°C and +40°C.
1
Valid from + 15°C to +30°C, <80% RH, for 1 year. Derate linearly to
2 times indicated values at + 5°C and +40°C.