43
Figure 4.20.
Connection ports on the back of the temperature control unit.
Technical specifications
Dimensions of control electronics box
130*230*185 mm3 (h*w*d)
Dimensions of sample compartment
40*82*82 mm3 (h*w*d)
Dimensions of sample compartment surrounded
with ceramic insulator
90*115*115 mm3 (h*w*d)
Heating plate power
150 W
Temperature range
from RT to 200
⁰
C (C203E) or 350
⁰
C (C203HT)
Accuracy
+/- 1oC
Voltage
Nominal/Operational
230 V, 50/60 Hz 115 V, 50/60 Hz
The whole sample compartment consists of the following items (Figure 4.21):
An insulating ceramic base plate with the heating plate and a cable for connecting to the heater port. The
bottom of the ceramic base plate is made so that it fits perfectly on the sample stage of Theta.
Environmental chamber (made of aluminium) with glass windows and a lid with a hole for the needle.
There is also one hole for the temperature sensor near one of the windows.
A insulating ceramic lid with a corresponding hole for the needle as in the environmental chamber, and
jackets for the windows.
Содержание Theta
Страница 1: ...Theta User manual OneAttension Optical Tensiometer Revision 1 0...
Страница 3: ...6 3 Theory of surface free energy 65 7 TECHNICAL SPECIFICATIONS 70 8 CONTACT INFORMATION 71...
Страница 41: ...38 Figure 4 15 Device controls and recipe sheet parameters for multiliquid dispenser...
Страница 54: ...51 Figure 4 30 Analysis tab of the PD 200 experiment Figure 4 31 Calculated results...