background image

AXIAL AND LATERAL RESOLUTION 

Description and Reason For Testing 

Resolution is the minimum reflector separation between two closely spaced objects which can be imaged 

separately. If a system has poor resolution capabilities, small structures lying close to each other will 

appear as one image, causing improper interpretation of the ultrasound findings. 

Resolution along the axis of the sound beam is referred to as Axial Resolution. Axial Resolution is 

affected by the pulsing section of the imaging system and the condition of the transducer. 

Resolution perpendicular to the axis of the sound beam is termed Lateral Resolution. Lateral Resolution 

is dependent upon the beam width, increased beam width will reduce the Lateral Resolution. 

Testing Procedure 

In the Model 535-H scan surface #1 and #3 are used to evaluate axial resolution at depths of 5.5 and 11 

cm. To evaluate lateral resolution, scan surfaces #2 and #4 are used, at depths of 4.5 and 16.5 cm. 

The interval spacing of the line targets are 5, 4, 3, 3, and 1 mm. Each target is spaced 1 mm laterally to 

avoid over shadowing of the adjacent targets.  

1.  Place the phantom on a clean, flat surface with scanning surface #1 positioned for use. 

2.  Apply an adequate amount of low viscosity gel or water to the scan surface. If water is used, fill the 

scanning well slowly to avoid introduction of air bubbles. 

3.  Adjust the instrument settings (TGC, output, etc.) to establish baseline values for "normal" liver 

scanning. If the bottom of the phantom is seen, adjust the gain settings until image goes entirely black. 

Record these settings on the quality assurance record. These settings should be used for subsequent 

testing. 

4.  Position the transducer over the axial resolution group of line targets on the phantom until a clear 

image is obtained. Freeze this image. 

5.  Examine the image to determine if all of the line targets within the group are clearly displayed as 

separate target points. Record the closest spaced target points which can be imaged (refer to 

specification drawing). Obtain a hard copy of the display. 

    

6.  Document all observations made on the quality assurance record. 

7.  Position the phantom with scan surface #3 ready for use. Repeat steps 2-5. 

8. Repeat steps 1-6 for scan surfaces #2 and #4 to obtain measurements of the lateral resolution. 

Results 

The system's ability to resolve the array targets at all four depths should remain consistent from week to 

week when using the same instrument settings and the Model 535-H phantom. Compare the test results 

obtained with a baseline or previous test. If the current image demonstrates changes in the system's 

ability to resolve these targets, corrective action should be considered. 

FOCAL ZONE 

Description and Reason For Testing 

The focal zone is the region surrounding the focal point in which the intensity and the lateral resolution is 

the greatest. Clinically, structures examined within the focal zone will provide the best diagnostic 

information obtainable. The focal zone can be affected by changes in the pulsing/receiving section of the 

imaging system or damage to the transducer. 

Содержание 535-H

Страница 1: ...MODEL 535 H BASIC QUALITY ASSURANCE PHANTOM Revised June 2014 ATS Laboratories Incorporated 900 Asbury Ave Norfolk VA 23513 Tel 800 617 1177 Email admin cirsinc com Website atslaboratories phantoms com ...

Страница 2: ...Diagram 5 Dead Zone 6 Vertical Measurement Calibration 6 Horizontal Measurement Calibration 7 Axial and Lateral Resolution 8 Focal Zone 8 Sensitivity Maximum Depth of Penetration 9 Functional Resolution Definition and Fill in 10 Care and Handling of Rubber Based Phantoms 11 Warranty 11 Sample Ultrasound Performance Record ...

Страница 3: ...To ensure measurement accuracy ATS incorporates a thermometer strip affixed to the outside surface of the phantom housing The sound velocity of most diagnostic imaging systems is calibrated to 1 540 meters per second mps the assumed average velocity of sound through human soft tissue The rubber based tissue mimicking material has a sound velocity of 1450 mps at 0 5dB cm Mhz at room temperature 23 ...

Страница 4: ...Scan Surface 1 Scan Surface 3 Scan Surface 2 Scan Surface 4 ...

Страница 5: ...umber of Targets 17 Interval Spacing 1 cm Depth 1 17 cm Horizontal Group Number of Targets 10 Number of Groups 2 Interval Spacing 2 cm Depth 5 cm from Scanning Surface 1 5 cm from Scanning Surface 3 Dead Zone Group Number of Targets 9 Lateral Displacement 5 mm Interval Spacing 1 mm Depth 2 10 mm Axial Lateral Resolution Group Number of Targets 6 Lateral Displacement 1 0 mm Interval Spacing 5 4 3 2...

Страница 6: ... depth of the dead zone 7 Document the depth measurement on the quality assurance record Results If the depth of the dead zone is greater than 10 0 mm corrective action should be considered by the individual Ultrasound Department VERTICAL MEASUREMENT CALIBRATION Description and Reason For Testing Vertical distance measurements are obtained along the axis of the sound beam Proper diagnosis is depen...

Страница 7: ...ration Linear array scanning systems should use 1 scanning surfaces Sector scanning systems should use 3 scanning surfaces 1 Place the phantom on a clean flat surface 2 Apply an adequate amount of low viscosity gel or water to the scan surface If water is used fill the scanning well slowly to avoid introduction of air bubbles 3 Adjust the instrument settings TGC output etc to establish baseline va...

Страница 8: ...l liver scanning If the bottom of the phantom is seen adjust the gain settings until image goes entirely black Record these settings on the quality assurance record These settings should be used for subsequent testing 4 Position the transducer over the axial resolution group of line targets on the phantom until a clear image is obtained Freeze this image 5 Examine the image to determine if all of ...

Страница 9: ...is the focal zone Measure the width of the beam and the depth at this point 7 Document the depth of the focal zone and the measurement of the focal width on the quality assurance record Results The location of the focal zone should agree with the manufacturer s specifications and should not change with time This applies to both fixed and dynamically focused systems If changes occur corrective acti...

Страница 10: ...ystem is capable of resolving at a given depth Definition and Fill in describes the imaging system s ability to detect and display the shape and echogenic characteristics of a structure Clinically a correct diagnosis is dependent upon the system s ability to differentiate between a cystic or solid structure versus echo patterns originating from the surrounding normal tissue Testing Procedure The M...

Страница 11: ...oided since they may adversely react with the rubber based material WARRANTY Statement of Warranty ATS Laboratories Incorporated warrants this rubber based phantom for it s lifetime from the date of delivery to the purchaser that the Phantom is free from functional defects in materials and workmanship The lifetime of this phantom is estimated to be 10 years from the date of manufacturing If ATS La...

Страница 12: ......

Отзывы: