8. All findings should be documented on the quality assurance record.
Results
The targets should appear circular with sharp clearly defined edges, indicating an abrupt transition from
the echogenic to the anechoic region. The targets are anechoic and should be free of any internal echoes
or fill-in. However, the presence of internal system noise may manifest itself by producing an observable
shade of gray within the target area.
The specific values determined, while significant in their own right, are somewhat less important than
stability over time. Performing this test on a routine basis at the same instrument settings should produce
the same results. Any changes should be investigated.
CARE RUBBER-BASED PHANTOMS
For best results the phantom should be kept clean at all times and stored at room temperature. In
particular a build-up of dried coupling gel on the scan surface should be avoided. The phantom may be
cleaned with warm water using a lint free cloth. Particularly stubborn stains and dirt may be removed with
a mild household cleaner. The use of petroleum solvents should be avoided since they may adversely
react with the rubber-based material.
WARRANTY
Statement of Warranty:
ATS Laboratories, Incorporated warrants this rubber-based phantom for it’s lifetime from the date of
delivery to the purchaser, that the Phantom is free from functional defects in materials and workmanship.
The lifetime of this phantom is estimated to be 10 years from the date of manufacturing. If ATS
Laboratories, Incorporated, deems the phantom to be defective, at its sole option, the Phantom will be
repaired or replaced free of charge, in a reasonable amount of time.
ATS shall not be otherwise liable for any damages, including but not limited to incidental damages,
consequential damages, or special damages.
There are no express or implied warranties which extend beyond the warranties as stated below.
Conditions of Warranty:
1. The defect must be reported and the Phantom returned within the warranty period.
2. The Phantom must be packaged properly to avoid damage during shipping.
3. All transportation charges will be paid by the purchaser.
Invalidation of Warranty:
1. If the phantom has been altered or repaired other than by ATS Laboratories, Incorporated.
2. If the phantom has been subject to abuse, misuse, negligence or accident.
3. If the purchaser has exposed the Phantom to petroleum solvents.
Содержание 535-H
Страница 4: ...Scan Surface 1 Scan Surface 3 Scan Surface 2 Scan Surface 4 ...
Страница 12: ......