APD series bench scale
APD_01_11.05_EN_U
12
-
Press the MODE/C key many times until the display shows the message “SAVE?”.
-
Press MR to confirm the changes made or another key for not saving.
The instrument stores the last selection made, also after it is turned off.
8.3 LIMITATION OF THE TARE FUNCTIONS FOR DIRECT SALE
With approved instrument, it is possible to limit the tare functions, by setting “yES” in the step
SEtuP >> d.SALE
(TECH.MAN.REF.)
.
In this case, all the tare operations will be disabled.
With approved instrument, the
d.SALE
step is read-only.
8.4 AUTO POWER OFF FUNCTION
It is possible to automatically turn off the scale (from 1 to 255 minutes), or disable it; the auto power off takes place when,
with unloaded scale
, the weight has not been moved or a key has not been pressed for the time set: the display shows the
“- OFF – “ blinking message and an acoustic signal is emitted; after this the scale turns off.
For the setting, follow the procedures below:
-
Turn on the scale, press TARE while the firmware version is displayed (the display shows the “F.ModE” menu).
-
Press MR to enter the menu.
-
Press ZERO many times (to scroll ahead through the parameters) or TARE (to scroll backwards) until one finds the
“En.SAVE” parameter.
-
Press MR to enter the menu
-
Press ZERO many times (to scroll ahead through the parameters) or TARE (to scroll backwards) until one finds the
“AutoFF” parameter.
-
Press MR to enter the parameter.
-
With the ZERO or TARE keys select the possible options: “diSAb” (auto switch-off disabled), “EnAb” (auto switch-off
enabled).
-
Confirm with MR; if “EnAb” has been selected, one will be asked to enter the number of minutes after which the scale
should turn off: enter a number between 1 and 255 (using the M+ key to select the digit to be modified and ZERO/TARE
to decrease/increase it) and confirm with MR.
-
Press many times the MODE/C key until the display shows “SAVE?”.
-
Press MR to confirm the changes made or another key for not saving.
8.5 LOW BATTERY WARNING
One should set the supply mode by setting the step
SEtuP
>>
ALiM
.
The charge level is shown in the weighing phase by the battery symbol:
-
:
battery is charged.
-
: battery is partially charged.
-
: battery is discharged: connect the scale to the mains for recharge the battery or replace it.
Furthermore, for a few seconds the "Low.bat " message appears on the display (minimum level voltage).
When recharging the battery, the display shows the recharging phase below:
RECHARGING PHASE:
…
RECHARGE IS COMPLETED:
NOTES:
- While recharging, the instrument can be used as usual.
- The instrument automatically turns off when the voltage goes below the minimum level.
- It’s possible to view the recharge percentile of the battery by pressing the
ZERO
key upon start-up (see the section
“POWER SUPPLY AND START-UP”).