11
DFT
Batch Template Packages
Templates for Creation Measurement Tasks
Industrial
The packages contain various batch types. That are batch templates with especially designed
measurement tasks for specific coating thickness measurement applications.
Template package contains following batch types:
•
Individual
Batch template for free configuration: All metrological standard functions are available
and calibration method Zero + 1 Foil is used
•
Elementary
Template with minimum configuration: No further metrological function settings (customiz-
able), only calibration method Zero is used
•
Smooth Surface
Preconfigured batch template for measurements on smooth surfaces: Settings of tolerance
limit values (customizable) and calibration method Zero + 1 Foil is used.
•
Rough Surface
Preconfigured batch template for measurements on rough surfaces: Display and storage
of the mean value of a given number (n) of measurements (customizable, single readings
are also stored) and the calibration method Zero + 2 Foils is used.
Corrosion
Template package containing batch types with especially designed measurement tasks for mea-
surements of corrosion-protective coatings. The template package contains following batch types:
•
Individual
Batch template for free configuration: All metrological standard functions are available
and calibration method Flexible is used
•
IMO PSPC
Preconfigured batch template containing 90/10 rule, calibration method and evaluation
for coating thickness measurement according to requirements of "Performance Standard
for Protective Coatings" of the International Maritime Organization (IMO PSPC), calibra-
tion method 2 Foils is used
•
SSPC PA2
Preconfigured batch template with settings (partly customizable), calibration method and
evaluation regulations for coating thickness measurement according to guideline SSPC-
PA2 of the Society for Protective Coatings (SSPC), calibration method 2 Foils is used
•
ISO 19840
Preconfigured batch template with settings (partly customizable), calibration method and
evaluation regulations for coating thickness measurement according to standard
ISO 19840, calibration method 2 Foils is used
•
AS 3894.3
Preconfigured batch template with settings (partly customizable), calibration method and
evaluation regulations for coating thickness measurement according to Australian stan-
dards AS 2331.1.4 and AS 3894.3-B, calibration method 1 Foil is used
•
SIS 184160
Preconfigured batch template with settings (partly customizable), calibration method and
evaluation regulations for coating thickness measurement according to Swedish standard
SIS 184160, calibration method 2 Foils is used
Metrological Standard Functions
Measurement Tasks
Batch
File containing all metrological function settings and the linking to calibration necessary for the
measurement task as well as the measured readings and evaluations
Block creation
Measured readings grouped in measurement blocks
Tolerance limits/Nominal thickness
Adjustable, depending on the selected batch type
Offset value/Correction value
Adjustable, is deducted automatically from the measured reading. Thus, one obtains the thickness
of the top coating if for instance the interim coating is known.
Measurement reading acquistion
Automatic upon placement of the gage probe