10.0 Technical data at complete equipment
Voltage
100 - 230 V~ ± 10 %; 50/60 Hz
Power consumption
Max. 1.0 A
Power consumption
Max. 200 VA
Fuses
2 x T 3.15 A/H 250 V
Other safety equipment
Internal fuses on control circuit board
LED light module
Illuminance: min. 195 kLux (in 5 cm distance of a 4.7 mm high-performance light
guide)
Colour temperature: 5.500 K ± 10 %
Removable adaptor
Olympus, Wolf, Pentax
LED power supply for ATMOS
®
LS
21 LED
700 mA regulated
Endoscope management
For rigid and flexible optics (max. length 500 mm), one quiver per clean and
contaminated optics.
Endsocope heating
For 4 quivers, quivers temperature approx. 40 °C
Inspection of disinfection time
Timer adjustable in 10 steps from 6 to 60 minutes
Camera module
see ATMOS
®
Cam 21 / 31
LED stroboscope
see ATMOS
®
Strobo 21 LED
Operating time
Continuous operation
Protective earth conductor
resistance
max. 0.1 Ω
Earth leakage current
max. 0.5 mA
Enclosure leakage current
max. 0.1 mA
Patient leakage current
max. 0.1 mA
Ambient conditions
Transport / storage
-10...+50 °C
30...95 % humidity without condensing
air pressure 500...1060 hPa
Operation
+10...+35 °C
30...95 % humidity without condensing
air pressure 700...1060 hPa
Dimensions H x W x D
88.5 x 41.2 x 54.0 cm
Weight
15 - 30 kg, depending on configuration
Period tests
Repeat test of the electrical safety every 12 months.
Recommended: inspection according to the manufacturer‘s specifications.
Safety class (EN 60601-1)
I
Type of protection
Application parts type BF
Type of protection
IPX0
Classification in accordance with
Annex IX to EC Directive 93/42/
EEC
Class I, Standard 12
CE marking
CE
UMDNS code
11-585
GMDN code
11585
Ident-Nr.
531.0000.0
Please note the technical specifications of the single devices in the separate operating instructions!
Technical data unchanged since 21.01.2016