9
• For cleaning, never use alcohol or alcoholic cleaning agents. Avoid any contact of the lens with alcohol!
• Please observe the operating instructions for use prescribed by the manufacturers of disinfectants. Pay attention regarding
concentration and material compatibility.
• Do not use:
- disinfectants which contain concentrated organic or inorganic acids as they could cause corrosion damage.
- Disinfectants containing chloramides, phenol derivatives or anionic tensides, as these may cause stress cracks in the
material used for the housing of the unit.
• Do not moisten the lens in the light object with liquids.
• The surfaces can be cleaned with a moist cloth.
• The device has to be cleaned and disinfected regularly, if it is applied by several users.
Surface disinfectants
Disinfectant
Ingredients
in 100 g Manufacturer
Green & Clean SK
Di alkyl dimethyl ammonium chloride
< 1 g Metasys, Rum (Austria)
Alkyl dimethyl ethyl benzyl ammonium
chloride
< 1 g
Alkyl dimethyl benzyl ammonium chloride
< 1 g
Dismozon
®
plus
(Granulate)
magnesium monoperoxyphthalate
hexahydrate
95.8 g Bode Chemie, Hamburg
5.0 Cleaning