Operating Instructions Xenotest
®
150 S+
• tight connection and possible leakage of water hoses,
• condition of the water filter,
• filling level of the instrument-integral water reservoir or of the supply reservoir
for treated water,
• condition of connector and power supply cable.
6.3
Installation and removing the xenon lamp
WARNING - Electric shock!
Contact with voltage-carrying components may cause a lethal
electric shock. The irradiator ignition system is under high voltage.
Prior to removing / installing the xenon lamps, disconnect the in-
strument from the power supply system!
• Switch the instrument off using the ON/OFF switch.
• Disconnect the mains plug from the socket and protect it from acci-
dental reconnection.
• Check to see if the instrument is de-energized.
CAUTION - Risk of burns!
The components of the irradiation unit reach extremely high tem-
peratures; contact with skin may cause burns.
Prior to replacing the xenon lamp, allow the irradiation unit com-
ponents to cool down sufficiently!
NOTE - Handling the xenon lamp:
Residues from finger contact may burn into the xenon lamp glass
tube and cause uneven irradiation.
• Grasp the xenon lamp only at the lamp socket.
• Wear clean cotton gloves for installation.
6
Start-up