Operating Instructions Xenotest
®
150 S+
25
5.5
Ventilation system
The Xenotest
®
150 S+ air-cooled. The cooling system is fed
by two separate air streams:
• Lamp cooling
• Test chamber cooling
Lamp cooling, Fig. 13:
The cooling air for the lamp cooling is drawn in by a fan th-
rough the vent slots
3
on the right side of the instrument.
The air stream flows through the assembly shaft on the right
side, cooling the electronic components first. The air is then
guided through the exterior cylinder
2
of the lamp unit to
cool the xenon lamp.
The heated air exits through the vent openings on the top of
the instrument
1
where it is either removed by a technical
ventilation system that prevents back-pressure or vented di-
rectly into the open.
Test chamber cooling, Fig. 14:
The cooling air
1
for test chamber cooling is drawn in th-
rough air inlet openings
7
by the fan
9
and flows through
the air filter
8
.
The fresh air intake volume is regulated by a motor-opera-
ted door
6
, depending on the test chamber temperature
setting.
The air stream is routed past the heating system
B
and he-
ated to the required temperature, after which it passes th-
rough the humidification system
C
where it is mixed with
humidified air in accordance with the predefined values be-
fore entering the test chamber
4
through the top of the
chamber.
The air stream is uniformly distributed over the sample ca-
rousel
3
while the temperature and relative humidity are
continuously measured by the integral sensor system
2
.
The used air
A
is exhausted through the air outlet opening
5
in the front left-hand area of the test chamber and, toge-
ther with the cooling air for lamp cooling, is either removed
by a technical ventilation system or vented directly into the
open.
5
Functional description
Fig. 13
Fig. 14