7
Maximum level sensor
LD-1
6.
MAINTENANCE
Every week:
1. Check the condition of the batteries. In the event of the green LED blinking once, it is recommended to replace the batteries
(3x1.5V).
Every six months:
1. Test the device by holding down the test button.
2. Check if the sensor tip is dirty. If dirty, clean it dry with water or a mild detergent (see the safety data sheet before cleaning).
3 Verify sensor performance by immersion test or rotating the sensor 180
˚
.