10

Lamp Installation or Replacement BIO-1.5 and BIO-3.0 using ballast no.10-

0518A 120v 50/60Hz and using ballast no.10-0514A 230v 50/60Hz

IN ORDER TO PERFORM THIS TASK, BE SURE TO WEAR THE FOLLOWING SAFETY 

EQUIPMENT: SAFETY GLASSES OR A FACE SHIELD, AS WELL AS GLOVES.

1. 

  

Disconnect power to water purifier, and remove ground connector from the chamber.

2. 

Remove 

EASY-OFF

TM

 End Cap by pulling cap off static gland nut. 

3. 

Pull lamp cord and carefully withdraw lamp approximately 2 inches from chamber. While holding lamp 

 

end, remove lamp socket from exposed lamp end.

4. 

Carefully withdraw lamp from chamber. Be sure to withdraw lamp straight out without angling 

 

until completely clear of quartz sleeve.

Lamp and quartz sleeve are easily damaged. Exercise care when handling.

5. 

Reinstall lamp in reverse order.

Germicidal ultraviolet rays are harmful to eyes and skin. Do not restore power to water purifier until 

lamp and both 

EASY-OFF

TM

 End Caps have been properly reinstalled.

Step 1 - Disconnect Power to Water 

Purifier and Remove Ground Connector

Step 4 - Remove Lamp

Step 3 - Remove Socket

Step 2 - Remove End-Cap

Figure 4 – Lamp Replacement for BIO-1.5 

 

and BIO-3.0 (ballast no 10-0518A or 10-0514A)

Содержание Bio-Logic BIO 1.5

Страница 1: ...Manufacturers Engineers Sales Service Germicidal Ultraviolet Equipment Lamps 375 Marcus Boulevard Hauppauge NY 11788 631 273 0500 Fax 631 273 0771 Email Sales AtlanticUV com Shop Products at AtlanticU...

Страница 2: ...Patent Notice 16 WARRANTY PRODUCT REGISTRATION 16 These instructions generally describe the installation operation and maintenance of the Bio Logic Ultraviolet Water Purifiers Models BIO 1 5 and BIO...

Страница 3: ...TLY hazardous situation which if not avoided WILL result in death or serious injury To guard against injury basic safety precautions should be observed including the following 1 Read and follow ALL sa...

Страница 4: ...ual indication of germicidal lamp operation Water leaving the purifier is instantly ready for use no further contact time is required Limitation of Use The water purifier is intended for the use with...

Страница 5: ...er should always be located closest to the point of use This reduces the chance of the purified water being re contaminated by bacteria in the water distribution system after the water purifier 6 As w...

Страница 6: ...6 Recommended Installation Figure 2 Recommended Installation...

Страница 7: ...ter be installed in line prior to the water purifier The sediment filter will stop or trap any particulates from entering the water purifier Particulates may cause damage to the quartz sleeve as well...

Страница 8: ...sensor is capable of operating an optional audio alarm and or solenoid valve 13 Promate Audio Alarm Activated by the Ultraviolet Monitor or Safety Sensor alerts the user to any malfunction detected 1...

Страница 9: ...actory with questions Inspection 1 Regularly inspect the water purifier to ensure that the germicidal lamp is still in operation An LED indicator light located on the ballast provides visual indicatio...

Страница 10: ...mber While holding lamp end remove lamp socket from exposed lamp end 4 Carefully withdraw lamp from chamber Be sure to withdraw lamp straight out without angling until completely clear of quartz sleev...

Страница 11: ...hin the static gland nut If so remove Teflon washer from static gland nut before proceeding 6 Carefully remove quartz sleeve from chamber NOTE It is advisable to support the quartz sleeve on the oppos...

Страница 12: ...o remove all quartz fragments from the interior of the chamber 8 Carefully discard all pieces of the broken quartz sleeve 9 Align the end of the quartz sleeve with the gland fitting of the chamber 10...

Страница 13: ...he Maintenance section Low input voltage Verify input voltage to purifier Change in water quality Have water tested to confirm that it does not exceed maximum recommended concentration levels for use...

Страница 14: ...special input power requirements Table 5 Optional Accessories Optional Accessories Available for SENTRY Safety Sensor BIO 1 5 BIO 3 0 Promate Time Delay Mechanism BIO 1 5 BIO 3 0 Promate Elapsed Time...

Страница 15: ...5 1366 R 05 1370 R 4 Static Gland Nut 2 25 1492D 25 1492D 5 Teflon Washer 2 25 3023A 25 3023A 6 Rubber O Ring Quartz Sleeve 2 00 1023A 00 1023A 7 CRYSTAL CLEAR Quartz Sleeve 1 15 1700A 15 1701A 8 Cham...

Страница 16: ...bility and fitness for a particular purpose whether or not purposes or specifications are described herein We further disclaim any responsibility whatsoever to the customer or to any person for injury...

Отзывы: