background image

 

Series A12

 

Unisens 

2-Wire Transmitter

 

 
 

O & M Manual 

- 4 - 

Rev-L, 7/15 

 

UniSens

  transmitters,  also  designated  Series  A12,  are  loop-powered  devices  capable  of 

operating  from  DC  power  supplies  between  12  and  30  VDC.    Normally,  this  transmitter  will  be  powered 
from the 24 VDC supply from a plant wide monitoring computer, a data logger, or an alarm systems that 
supplies the DC voltage.  The output of the transmitter is a 4-20 mA signal which is linear with respect to 
concentration over the range of the sensing module. 
 
 

Construction of the 

UniSens

 transmitter is modular to allow simple installation and maintenance 

as shown in Figure 1 on the next page.  The transmitter electronics is housed in an aluminum shell with 
an 11 pin relay plug on the bottom.  The outer housing is an explosion-proof and weatherproof  enclosure 
with  a  relay  base  mounted  inside.    The  transmitter  electronics  module  simply  plugs  into  the  relay  base, 
and  all  wiring  connections  are  made  to  that  base.    Figure  1  shows  an  exploded  view  of  a  complete 

UniSens

 transmitter with sensing module. 

 
 

The  sensing  elements  used  in  the 

UniSens

  are  electrochemical  sensors  manufactured  by  ATI.  

These sensors are designed for use in ambient air or in ventilation ducts at temperatures of -30° to +50° 
C (some sensors rated to -40° C, oxygen sensors rated to -10° C) and at relative humidities between 20 
and  98%  RH.    Use  of  these  sensors  in  extremely  dry  air  or  in  condensing  gas  streams  is  not 
recommended.   
 
 

Each  sensing  element  is  incorporated  into  a  snap-in  sensing  module,  which  also  contains 

electronic memory that is read by the transmitter.  This memory contains information on what type of gas 
sensor is contained in the  module, the range of that  particular module, and  the  calibration constants for 
the  sensor.    When  the  module  is  snapped  into  a  transmitter,  this  data  is  read  into  the  transmitter 
electronics  and  information  related  to  the  module  is  displayed  on  the  LCD  display  on  the  transmitter.  
Because all calibration data is stored in the sensing module, these modules can be calibrated using any 
spare  transmitter  and  simply  snapped  into  any  other  transmitter  for  operation.    This  unique  feature 
eliminates the need for field calibration. 

 

AUTO-TEST OPTION 

 
 
 

UniSens

  transmitters  are  available  with  an  optional  feature  that  eliminates  much  of  the  manual 

response testing required for toxic  gas detection systems.  This feature  is referred to as  Auto-Test, and 
consists  of  a  miniature  electrochemical  gas  generator  close  coupled  to  the  sensor.    The  generator 
produces  a  "puff"  of  gas  when  activated  by  the  transmitter.    The  response  of  the  sensor  is  monitored 
during this test to be sure that it is still responding to the particular gas, and the output of the transmitter is 
driven  below  3.7  mA  if  the  sensor  does  not  respond.    At  the  same  time,  the  LCD  display  indicates  that 
there was an Auto-Test failure. 
 
 

The Auto-Test sensor check occurs automatically every 24 hours so that the sensor response is 

verified daily.  In addition, the Auto-Test can be activated manually at any time using the controls on the 
front of the transmitter.  During the Auto-Test cycle, the 4-20 mA output is held at 4 mA so that receiving 
devices do not alarm during the test.  Because most users manually test toxic gas sensor response once 
a week, the Auto-Test feature can pay for itself very quickly in reduced labor cost. 
 

 

 

Содержание UniSens A12 Series

Страница 1: ...logy Inc ATI UK Limited 6 Iron Bridge Drive Unit 1 2 Gatehead Business Park Collegeville PA 19426 Delph New Road Delph Phone 800 959 0299 Saddleworth OL3 5DE 610 917 0991 Phone 44 0 1457 873 318 Fax 6...

Страница 2: ...TION 20 CALIBRATION 21 SENSING MODULE ZERO ADJUSTMENT 22 SENSING MODULE SPAN ADJUSTMENT 23 CALIBRATION OF DUCT MOUNT SENSORS 24 MA OUTPUT ADJUSTMENT 25 UNDER RANGE ENABLE DISABLE SELECTION 25 AUTO TES...

Страница 3: ...18 0 4 0 2000 PPM 0 10 Nitric Oxide 25 0 50 PPM 0 25 PPM 0 500 PPM Oxygen 19 0 25 0 10 0 100 Phosgene 20 0 2 PPM 0 2 PPM 0 200 PPM OXIDANT GASES Bromine 10 0 2 PPM 0 2 PPM 0 100 PPM Chlorine 11 0 10 P...

Страница 4: ...ransmitter This memory contains information on what type of gas sensor is contained in the module the range of that particular module and the calibration constants for the sensor When the module is sn...

Страница 5: ...Series A12 Unisens 2 Wire Transmitter O M Manual 5 Rev L 7 15 Figure 1 UniSens Transmitter Components...

Страница 6: ...spaces with easy removal of the sensor for service The special mounting fitting provides a method for feeding calibration gas to the sensor without complete removal from the duct Duct mount sensors a...

Страница 7: ...ft Application dependent but generally less than 3 per month Under Range Trip 20 when enabled Output Loop powered 4 20 mA 750 ohms maximum at 24 VDC loop power Power 12 30 VDC Enclosure NEMA 4X and Ex...

Страница 8: ...required by the local electrical code The conduit or cable gland entry into the enclosure must be sealed with Dow Corning 732 multipurpose sealant The sealant should be liberally applied to the 3 4 1...

Страница 9: ...conduit fasteners are used A mounting bracket is supplied with the transmitter for fastening the transmitter enclosure to a wall or mounting plate Figure 4 shows the hole center dimension for that mou...

Страница 10: ...e 00 0814 adapter be screwed into the customer supplied adapter as shown in Figure 5 below Screw in the adapter so that the calibration gas inlet port is oriented in a direction that gas tubing can ea...

Страница 11: ...30 second time delay on this trouble indication When a sensing module is removed the LCD display will begin counting down from 30 During the countdown the output will be held it s value prior to unpl...

Страница 12: ...th ground connection point The transmitter module can be removed from the outer enclosure by gently rocking it back and forth while pulling out Sensing module holders are available with up to 5 feet o...

Страница 13: ...Series A12 Unisens 2 Wire Transmitter O M Manual 13 Rev L 7 15 Figure 7 Remote Electrical Connections...

Страница 14: ...sing module revision level XX Two digit gas code number see table 1 XXXX PPM Number indicating the full scale range of the sensing module Ur E Ur d Ur E Indicates that the signal trouble on under rang...

Страница 15: ...period the 5 minute start up delay may not be sufficient for the sensor to completely zero Some sensors can take up to 30 minutes to completely stabilize when unpowered for an extended period of time...

Страница 16: ...sor module or adjusting the 20 mA output value D Down Key Indicator activated when the magnetic control marked is activated E Up Key Indicator activated when the magnetic control marked is activated F...

Страница 17: ...f menu selections Through menu selections the user can review information about the transmitter calibrate the transmitter manually activate the Auto Test function if installed simulate 4 different mA...

Страница 18: ...Series A12 Unisens 2 Wire Transmitter O M Manual 18 Rev L 7 15...

Страница 19: ...N MODE To review the transmitter information activate the MODE control until the display shows InFo and then activate the ENTER control The display will scroll through the same information shown durin...

Страница 20: ...ity to simulate 4 different current output values in order to verify complete loop integrity and to calibrate receiving devices The output may be set to values of 4 0 12 0 and 20 0 and may also be set...

Страница 21: ...alibration gas sources Calibration span gas for some gases such as HCl Ozone or Hydrides are either very expensive or not available at all Calibration of these sensors require specialized equipment an...

Страница 22: ...available as part of all ATI calibration kits or may be obtained from any specialty gas supplier When zero air is to be used a calibration adapter part no 00 0248 must be used The calibration adapter...

Страница 23: ...liable span gas If possible a span gas with a gas concentration of at least 25 of the full scale sensing module span is recommended For example if the full scale span is 20 PPM span gas of at least 5...

Страница 24: ...nnection on the side See figure 11 below for the sensor orientation when in the calibrate position To zero and calibrate the duct sensor pull the sensor out of the duct until the retaining pin clicks...

Страница 25: ...his function will be enabled so that sensor under range conditions greater than 20 full scale will cause the transmitter to output the trouble value of 3mA To disable this function start from the norm...

Страница 26: ...will still perform its function Activating the ENTER key will clear the alarm Should the sensor detect a level above 50 of the full scale range the failure condition will be over ridden and the output...

Страница 27: ...ule keeper 05 0017 Kit Wall Mounting Bracket SENSING MODULES range for each module must be specified No A T D10 With Auto Test D28 00 0211 00 0212 Bromine 00 0213 00 0214 Chlorine 00 0215 00 0216 Chlo...

Страница 28: ...E10 28 Low Range Hydride 0 1000 PPB Gas Sensor 00 0279 E10 20 Phosgene Gas Sensor 00 0280 E10 34 Hydrogen Peroxide Gas Sensor 00 0292 E10 14 Low Range Ozone Gas Sensor 00 0681 E10 32 Super High Range...

Страница 29: ...0 Phosgene 00 0822 G10 21 Hydrogen Chloride 00 0823 G10 22 Hydrogen Cyanide 00 0824 G10 23 Hydrogen Fluoride 00 0825 G10 24 Hydrogen Sulfide 00 0826 G10 25 Nitric Oxide 00 0827 G10 26 Nitrogen Dioxide...

Страница 30: ...exposure to excessive gas concentrations Should inspection indicate that sensors have been expended rather than failed prematurely the warranty shall not apply The Manufacturer assumes no liability f...

Страница 31: ...DUCTS NH3 Ammonia CO Carbon Monoxide H2 Hydrogen NO Nitric Oxide O2 Oxygen CO Cl2 Phosgene Br2 Bromine Cl2 Chlorine ClO2 Chlorine Dioxide F2 Fluorine I2 Iodine HX Acid Gases C2H4O Ethylene Oxide C2H6O...

Отзывы: