
ATI
Q46P pH System
Part 6 – Calibration
49
O&M Manual
Rev-F (8/17)
flash. Use arrow keys to modify value; range is 0 to 999 seconds. Press
ENTER to store the new value.
*A Phasing
This function establishes the direction of the relay trip. When phase is
HI, the relay operates in a direct mode. Therefore, the relay energizes
and the LCD indicator illuminates when the pH value
exceeds
the
setpoint. When the phase is LO, the relay energizes and the LCD
indicator illuminates when the pH level drops
below
the setpoint. The
failsafe setting does have an impact on this logic. The description here
assumes the failsafe setting is OFF. Press ENTER to initiate user entry
mode, and the entire value will flash. Use the UP arrow key to modify
the desired value; selections include
HI
for direct operation or
LO
for
reverse operation. Press ENTER to store the new value.
See Figure 33 below for a visual description of a typical control relay application.
Figure 33 - Control Relay Example, Hysteresis and Opposite Phase
If Relay A Mode is set to Alarm Mode,
AL
, then the following settings will
appear in the Config Menu list automatically. In this mode, two setpoints
can be selected on the same relay, to create an alarm band. Phase HI
selection causes the relay to energize outside of the band, and Phase
LO causes the relay to energize inside of the band. This feature enables
one relay to be used as a control relay while the other is used as a HI-LO
Alarm relay at the same time. Setpoint A-LO must be set lower than
Setpoint A-HI. When AL mode is first selected, Setpoint A-LO is
defaulted to 0.
Figure 34 is a visual description of a typical alarm relay application.
When value rises to
≥
9.00 pH, relay closes.
When value falls to
≤
8.95 pH, relay opens.
*Setpnt A-HI
*Hyst A-HI
*Delay A-HI
*Setpnt A-LO
*Hyst A-LO
*Delay A-LO
Setpoint: 9.00 pH
Hyst:
0.05
Delay:
000
Failsafe: OFF
When value rises to
≥
9.05 pH, relay opens.
When value falls to
≤
9.00 pH, relay closes.
9.00 pH
PHASE: HI
ON
HYSTERESIS
OR
“DEADBAND”
X
OFF
8.95 pH
9.05 pH
PHASE: LO
HYSTERESIS
OR
“DEADBAND”
X
ON
9.00 pH