ATI
Q46H/63 Combined Chlorine System
Part 1 - Introduction
7
O&M Manual
Rev-K (11/18)
Figure 3 below shows an installation using a 00-1522 sealed flowcell for the chlorine sensor and
a 00-1527 sealed flowcell for the pH sensor. This type of installation requires careful flow control.
We recommend the use of our 03-0372 flow control assembly when using sealed flowcells. This
assembly consists of an in-line filter and a fixed-flow regulator which will maintain a constant 400
cc/min flow rate through the system. This flow will be maintained so long as inlet pressures are
between 5 and 125 PSIG. The in-line filter is used mainly to protect the flow control element
against larger particles that might cause plugging of the device.
Figure 3 - Sealed Flowcell Assemblies with Flow Control