Q45D Portable Dissolved Oxygen
Part 8
– System Maintenance
7.4
Lead AnodeReplacement
Galvanic D.O. sensors consume the lead electrode during normal operation. As oxygen
is measured, lead is converted to lead oxide, and after a period of time, the lead is
expended. The lead electrode in ATI's D.O. sensor cartridge can be easily replaced, and
replacement should be done automatically every 12 months.
The lead electrode is the thick lead wire wrapped around the sensor body see Figure 6.
It is connected through a stainless steel post with a nut and star washer. To change the
lead electrode, remove the nut and washer and unwrap the old lead. Loop the end of a
new lead electrode around the post and replace the nut and washer. Tighten the nut
firmly but do not over tighten as damage to the sensing module can result. Wrap the
remainder of the lead around the sensor body. There is no need to secure the other end
of the lead.