Manual, F/T Sensor, Data Acquisition (DAQ) Systems
Document #9620-05-DAQ.indd-20
Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 • Tel: 919.772.0115 • Fax: 919.772.8259 •
• Email:
57
5.5 Environmental
The standard F/T system is designed to be used in standard laboratory or light-manufacturing conditions.
Transducers with an IP60 designation are able to withstand dusty environments. Transducers with an IP65
designation are able to withstand dusty environments and wash down, and those with an IP68 designation
are able to withstand dusty environments and freshwater immersion to a specified depth. See ATI Industrial
Automation manual (9620-05-Transducer Section—Installation, Operation and Specification Manual) for
transducer environmental information.
IP60 and non-IP rated transducers, the PS box, and the IFPS box can be used in environments with up to
95% relative humidity, non-condensing.
Table 5.1—Interface Box Temperature Ranges
Model
Storage
Operation
Units
PS box
-30 to 75
0 to 60
°C
IFPS box
-30 to 75
0 to 60
°C
Note: These temperature ranges specify the storage and operation ranges in which the electronics can
survive without damage. They do not take accuracy into account.