32
Request Data Fields:
A 10 character ASCII representation
of a numeric value specifying what
to calibrate.
0 - Thermocouple
1 - RTD, Resistive Thermal Device
2 - Linear
3 - RAS, Remote Analog Setpoint
Response Data Field:
Ignored.
Command:
Perform Process High Calibration
Parameter #:
03
Description:
Performs a High Calibration. The data
field in the request message specifies
the process. Make sure the prerequi-
site for the calibration is satisfied
before issuing a calibration command
For instance, the RTD calibration
command must only be used when the
input sensor type is chosen as RTD or
RTD w/ Decimal.
Request Data Field:
A 10 character ASCII representation
of a numeric value specifying what
to calibrate.
Digital
Communications
Option
Содержание 16C series
Страница 1: ...SERIES 16C TEMPERATURE PROCESS CONTROLLER Standard Options and Digital Communications User s Guide ...
Страница 2: ... Copyright 2004 Athena Controls Inc ...
Страница 14: ...11 Series 16C with Digital Communications Option ...
Страница 38: ...35 Communications Option Menu Option ED SPI Protocol ...
Страница 39: ...NOTES ...