Druck PV 411 User Manual
Page 17
K258 Issue No. 1
Fault Finding
If system pressure reduces, check the following:
1.
Check the selector valve is in the correct position and properly
tightened.
2.
Allow sufficient time after pressure generation for the
temperature to stabilize.
3.
Check the adaptors, flexible pipe and connections between
the pump and equipment under test for leaks. Tighten any loose
joints and replace any seals that are worn or damaged.
4.
In hydraulic mode, if the volume adjuster can be wound fully
in, but maximum pressure cannot be achieved, there is probably air
trapped in the system. Re-prime, and repeat.
If, for any reason, a fault occurs within the pump, it is
recommended that the equipment be returned to an appointed
agent.