background image

 

 

 

 

2.  Locating a hot spot aim the thermometer outside the 

area of interest, then scan across with up and down 
motions until you locate the hot spot.(please turn on the 
laser to for accurate measuring) 

 

 

 

 

 

 

 

 

 

 

3. Diagram 

description 

(1)

 

Trigger

Press for turn on, and then display test result 

and hold data 7 seconds automatically ("HOLD") after 
unclamping switch. Turn off automatically after 10 
seconds without operate.

 

(2) 

Laser pointer button 

press it for turn on laser 

pointer, press again for turn off. 

(3) 

Back light /Up button 

In “Alarm temperature” 

and “setting emissivity” mode for adjust value up. In 
other mode for turn on back light. When product 
working, press it for turn on back light, press again for 
turn off.   

(4) 

MODE button

Press MODE button for cycle options 

- MAX - AVG - MIN –HAL –LAL mode   
A

MAX

Measure maximum data current

 

B

MIN

Measure minimum data current

;;

 

C

AVG

Calculate the average of all measure data

 

D

HAL

Alarm of high temperature. Option HAL mode, 

press

UP or DN

button for set alarm temperature. 

When LCD display ”H” means measure result exceeded 
the alarm temperature

 

E

LAL

Alarm of low temperature. Option HAL mode, 

press

UP or DN

button for set alarm temperature. 

When LCD display ”L” means measure result under the 
alarm temperature

 

(5) 

T/ DN button

 

In “Alarm temperature” and “setting 

emissivity” mode for adjust value down . When turn on 
product and press direct for 

/   select. 

(6) 

SET button

Press for setting emissivity. And press 

“UP or DN” for adjust from 0.1~1.0. Press SET button 
again for exit setting mode. 

(7)

 

LCD display

 

(8) 

Battery door

When replace battery, please using 

the finger indents to pull open the battery door. 

 

3. Maintenance 

1)  Lens cleaning: Blow off lose particles using clean 

compressed air. Gently brush remaining debris away 
with a moist cotton cloth. 

2) Case cleaning: Clean the case with a damp 

sponge/cloth and mild soap. 

3)  Please take out the battery when not using for a long 

time. 

Note: 
1)  Do not use solvent to clean lens. 
2)  Do not submerge the unit in water. 

 

4. Specifications 

Temperature

range 

Infrared

-30

 to 550

(-22 to1022

Accuracy 

±

3

or 

±

3%of rdg,-30

 to 0

(-22 to 32 ) 

±

2

or

±

2% of rdg,0

 to 100

(32 to 212

±

3

or 

±

3% of rdg,

100

(212

Display spot size 

10:1 

Repeatability 

1%of reading or 

±

1

 

Response time 

500ms 

Spectral response 

8-14um 

Emissivity 0.1

1.0 adjustable 

Relative humidity 

10-95% RH non-condensing 

Operating 
temperature 

0

 to ~60

(32 to 140

Storage 
temperature 

-20~60

(-4~140

) without battery 

Power Supply 

9V 6F22 battery 

Battery life 

Laser models:12 hrs 

Product size 

165

×

80

×

41mm 

Product weight 

155g

without battery

 

Note: 
Display “AL” means ambient temp lower than 0 ”; 
Display “AH” means ambient temp higher than 60

”; 

Display above code in normal ambient temp, probable 
means this meter was broken. 

 
Attached list 

Applicable Emissivity for 

Different Material ( For reference only) 

Material Emissivity 

Material 

Emissivity 

Aluminum 0.30 

Iron 

0.70 

Asbestos 0.95 

Lead 

0.50 

Asphalt 0.95 

Limestone 0.98 

Basalt 0.70 

Oil 

0.94 

Brass 0.50 

Paint 

0.93 

Brick 0.90 

Paper  0.95 

Carbon 0.85 

Plastic 

0.95 

Ceramic 0.95 

Rubber 

0.95 

Concrete 0.95 

Sand 

0.90 

Copper 0.95 

Skin 

0.98 

Dirt 0.94 

Snow 0.90 

Frozen food 

0.90 

Steel 

0.80 

Hot Food 

0.93 

Textiles 

0.94 

Glass(plate) 0.85 

Water 

0.93 

Ice 0.98 

Wood 0.94 

 

 

Отзывы: