In this mode is instrument anytime after switching ON. It is possible to enter this mode from 

other modes by pressing or by repeating pressing of MENU key. If the probe is not connected 
properly or temperature is out of the range symbol “--“ is displayed. 
 

 
Temperature is displayed on the upper LCD line (°C EXT) and relative 
humidity in %RH is displayed on the lower LCD line. Those readings are 
from the external probe.  
 
 
 
Press DISPLAY key to display other readings - ambient temperature from 
internal sensor (°C INT) and the dew-point temperature calculated from 
external probe values (symbol DP). 

 

 
Press DISPLAY key to display other readings - atmospheric pressure in 
hPA on upper line and dew point temperature on lower line (°C DP). 
 
 
 
Press DISPLAY key to display pressure tendency (hPa), which is the 
difference  between actual pressure and pressure before 3 hours. If the 
instrument is not switched ON at least 3 hours, pressure tendency is not 
available yet and symbol „--„ is display. 

 
 
 
Function HOLD (storing of actual measured values) and minimum a maximum memory 
 

Press HOLD  key in the default mode (displaying of actual measured value) to store actual 

measured values to internal memory (indicated by short beep). Anytime it is possible to display 
stored values from MENU (see below). Each pressing of the HOLD key in the default mode 
causes  values stored in HOLD memory are replaced with actual ones. 

 

Switched ON instrument permanently updates minimum and maximum  
memory of each measured values. Press MIN key (resp. MAX key) in the 
default mode to display minimum (resp. maximum) reading. These 
minimum and maximum readings are indicated by MIN (MAX) symbols 
on the LCD. Press DISPLAY key to display minimum (resp. maximum) 
value of other values. Pressing MIN (MAX) or MENU key again to return 
to default mode. Minimum and maximum memory is cleared from menu 
after confirmation selection CLR (see below). Values in HOLD, MIN and 
MAX memories remain stored even after instrument is switched OFF. 

 
 
Functions and settings available from menu 
 

Press MENU key to enter mode of viewing menu items one by one. Press arrow keys up and 

down to list all menu items. Press MENU key again to return to default mode (displaying of actual 
measured values). 

Содержание COMMETER C4141

Страница 1: ...COMMETER C4141 Digital thermo hygro barometer with external probe Instruction manual ...

Страница 2: ... 0 4 C Humidity Range measurement 0 to 100 RH Resolution 0 1 RH Accuracy 2 5 RH at the range of 5 to 95 RH Dew point temperature calculated from temperature and humidity Range 50 to 60 C Resolution 0 1 C Accuracy 0 5 C at the range of 30 to 95 RH Atmospheric pressure Range of measurement 800 to 1100hPa Resolution 0 1 hPa Accuracy 2hPa at ambient temperature of 23 C Measuring interval and display r...

Страница 3: ...internal 9V battery is replaced with rechargeable NiCd accumulator In usual operation from adapter accumulator is charged only with small current If accumulator is totally discharged its full charging in instrument takes approximately 100 hours Instrument with accumulator is not recommended for permanent operation without ac dc adapter plugged Accumulator works only as a standby source in case of ...

Страница 4: ...actual measured values and minimum a maximum memory Press HOLD key in the default mode displaying of actual measured value to store actual measured values to internal memory indicated by short beep Anytime it is possible to display stored values from MENU see below Each pressing of the HOLD key in the default mode causes values stored in HOLD memory are replaced with actual ones Switched ON instru...

Страница 5: ...partially loaded battery is displayed This value illustrates battery condition This selection by pressing ENTER key enables to enter mode of pressure correction to the sea level see below Display refresh mode is indicated In the FAST mode refreshment is fastest with regular interval approximately 0 7 s In the dynamic refresh mode DYN each 15 s refresh interval of display is doubled to maximum 5 s ...

Страница 6: ...measurement display starts to display the value out of limits If at least one alarm is active audio indication can be activated if menu AUDI On is selected see setting described above Alarm activation of each value can be disabled by setting lower alarm limit of the desired value up to its maximum This is indicated by OFF reading at the position of numeric value Value of upper limit of the same al...

Страница 7: ...cessary to replace battery with instrument switched OFF otherwise setting of d REF and AUDI from menu selections and data in memory HOLD MIN and MAX will be lost For the same reason do not disconnect the battery for longer than 1 minute even if instrument is switched OFF If it happens or if battery is totally discharged it is necessary to set again in appropriate menu selection LCD refreshment mod...

Отзывы: