Atal ATC-06 Скачать руководство пользователя страница 2

2

Manual for use of thermometer-hygrometer ATAL ATC06

Instrument  is  designed  for  measurement  of  temperature  and  relative  humidity  with  the 

possibility of displaying the dew point temperature. Measured values are displayed on a dual line 
LCD  display.  Temperature  is  measured  by  RTD  sensor  Ni1000/6180ppm.  Instrument  compares 
measured  values  of  temperature,  humidity  and  dew  point  with  two  adjustable  levels  for  each 
measured quantity. Breaking the level is indicated by  blinking the proper value on display and by 
audio indication (switchable). Instrument is equipped with minimum and maximum memory and 
Hold function. Minimum and maximum values and Hold value are possible to display on the LCD 
anytime.

Technical parameters:

Parameters of measurement:
Temperature: Range of measurement: -30 to +60 °C

Resolution: 0.1 °C
Accuracy: ±0.4 °C

Humidity:

Range measurement: 5 to 95 %RH
Resolution: 0.1 %RH 
Accuracy: ±2.5 % RH at the range of 5 to 95 %RH

Dew point temperature (calculated from temperature and humidity):

Range: -50 to +60 °C
Resolution: 0.1 °C 
Accuracy: ±0.5 °C at the range of 30 to 95 % RH

Measuring interval and display reading refresh: approximately 0.7 s in FAST mode

       

           0.7 to 5 s in dynamic mode

Power: battery 9V or ac/dc adapter 12V with NiCd accumulator 9V 
Average current consumption: 0.15 to 0.7 mA (depending on operation mode) 
Typical life of Zinc-Chloride battery: 4 months 
Typical life of Alkali-Mangan battery: 6 months

Operation conditions:
  Ambient temperature range: -10 to +60 °C
  Ambient relative humidity range: 5 to 95 %RH
  Outer characteristics in accordance with EN 33-2000-3: normal environment with 

characteristics AD1, AE1, AF1, AG1, AH1, AK1, AL1, AN1, AP1, AQ1, AR1, AS1, 
BA1, BE1

Not allowed manipulation: it is not allowed to touch sensors under the cover to avoid 
  sensors damaging or to effect calibration.
The sensors (under the cover) should not be exposed to direct contact with water or other 
liquids.
Storing conditions: temperature -10 to +60 °C relative humidity 5 to 95 %RH 
Dimensions: 141 x 71 x 27 mm 
Weight including battery: approximately 150 g

Battery  life  depends  on  selected  display  refresh  mode  (see  below).  In  FAST  mode  display  is 

refreshed in shortest possible interval with highest current consumption. In dynamic mode display 
is  refreshed  in  interval  up  to  5  s  in  case  measured  values  remain  stable.  Refresh  interval  is 
shortened  to  approximately 0.7  s  only  if  measured  values  change.  Current  consumption  in  this 
mode  in  usual  operation  is  lower,  battery  life  is  up  to  4  times  longer.  The  FAST  mode  is 
recommend to use only in cases, when slower display response is not acceptable.

Содержание ATC-06

Страница 1: ...ATAL ATC 06 Digital thermometer hygrometer Instruction manual ...

Страница 2: ...ttery 9V or ac dc adapter 12V with NiCd accumulator 9V Average current consumption 0 15 to 0 7 mA depending on operation mode Typical life of Zinc Chloride battery 4 months Typical life of Alkali Mangan battery 6 months Operation conditions Ambient temperature range 10 to 60 C Ambient relative humidity range 5 to 95 RH Outer characteristics in accordance with EN 33 2000 3 normal environment with c...

Страница 3: ...nt anytime after switching ON It is possible to enter this mode from other modes by pressing or by repeating pressing of MENU key Temperature in C is displayed on the upper LCD line and relative humidity in RH is displayed on the lower LCD line Press DISPLAY key to display other readings temperature in C on upper line and dew point temperature on lower line C DP Function HOLD storing of actual mea...

Страница 4: ...NTER key to display values stored in the HOLD memory Press DISPLAY key to display other stored values dew point temperature Press MENU key to leave this mode Battery voltage of partially loaded battery is displayed This value illustrates battery condition Display refresh mode is indicated In the FAST mode refreshment is fastest with regular interval approximately 0 7 s In the dynamic refresh mode ...

Страница 5: ... the arrow key UP to make value increase fast Press and hold the arrow key DOWN to make value decrease fast Release the arrow key and press ENTER to confirm new limit Reading ALHI indicates adjusted upper limit of the same alarm here air temperature Set the desired value in the same way as in above lower limit If needed it is possible to get back to lower limit setting of the same alarm by pressin...

Страница 6: ...battery for longer than 1 minute even if instrument is switched OFF If it happens or if battery is totally discharged it is necessary to set again in appropriate menu selection LCD refreshment mode d REF alarm audio indication AUDI and clear the minimum and maximum memory CLR ATAL instruments passed the following electromagnetic compatibility EMC tests radiation EN 55022 class B immunity EN 61000 ...

Отзывы: