background image

Zero-setting and Measurement 

 

Zero-setting 

【Caution】 

 

Zero-set the instrument at the beginning of each day before use as well as after replacing the batteries. 

 

Let water on the prism acclimate to the temperature of the instrument before zero-setting. 

 

When 

AAA

 is displayed, wipe the prism clean, apply water, and press the ZERO button again.

 

 

1.

 

Wipe the prism clean and 

apply about 0.3mL of tap water. 
 

 

 

2.

 

Press the START 

button, and after 

---

 

appears, the 
measurement of the 
main scale and the prism 
temperature are 
displayed. 

 

 

3.

 

 

 

4.

 

Zero-setting is 

complete when 

000

 blinks 

twice and then stops. 

 

Measurement : not 0 

Press the ZERO 
button while the 
water is on the 
prism. 

 

 

Measurement : 0 

Proceed to measurement. 
Wipe the prism clean with 
tissue paper. 

The instrument is zero-set 
correctly. 

 

Measurement 

【Caution】 

 

Do not use metal tools to apply samples on the prism as they may scratch the prism. 

 

Initial measurements may fluctuate with hot or cold samples. Wait for the instrument to acclimate to the sample temperature, 
approximately 20 seconds, to press the START button. Alternatively, press the START button multiple times until 
measurements become stable.

 

 

Do not splash water above 50℃. The plastic may warp, which may compromise the water resistance. 

 

When measuring hot samples, place only the necessary amount and do not let it overflow from the sample stage well. 

 

When hot water is necessary to clean off hardened samples, use water-soaked gauze around the prism area and keep hot 
water away from the body case.   

 

The displayed temperature is that of the prism and may not necessarily match the temperature of the sample. 

 

1.

 

Wipe the prism clean and 

apply about 0.3mL of the sample. 

 

 

2.

 

Press the START 

button 
 

 

 

 

3.

 

After 

---

 appears, the 

measurement of the main 
scale and the prism 
temperature are displayed. 
 

 

 

 

Display method for 

secondary scale 

Press the START button 
for 2 seconds while 
showing main scale. 
 

 

 

<To display the main scale> 

Press the START button and measure again. 
 

<LCD Auto Shut-off> 

The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off, hold down the START button for more than 
4 seconds. 
 

<For oily/fatty samples> 

Try stirring the sample on the prism while measuring to improve the 
repeatability of oily/fatty samples. 
 

 

 

Cleaning 

 

Caution

】 

 

Do not scratch the prism. 

 

The instrument is water-resistant, not waterproof, and should not be submerged. 

 
1. Wipe off the sample. 
2. Clean the prism and sample stage using a mild soap and thoroughly rinse with water. 
3. Dry the area with tissues thoroughly.

 

 

Error Messages 

 

The following messages alert the user when an operation has failed. 

 

The battery is low. 

   

The prism temperature is below the temperature 
range.   

 

The ZERO button was pressed with nothing or 
something other than water on the prism. 

 

 

The prism temperature is above the temperature 
range. 

 

The START button was pressed with nothing or an 
insufficient amount of sample on the prism. 

 

 

The instrument is faulty.   
(Replace the batteries. Contact ATAGO if this 
error persists.) 

 

The sample measured outside the measurement 
range. 

   

 

Too much light is entering the prism, and the 
instrument cannot measure accurately. 
(Shade the sample stage with your hand and take 
a measurement again.) 

   

 

 

Replacing the Batteries 

 

【Caution】 

 

Fasten the battery compartment cover tightly to prevent water ingress or poor connection, which will
cause erroneous measurements. Push the cover in firmly and turn.   

 

When the O-ring on the cover is dirty or damaged, the water resistance may be compromised. 
Lubricate the O-ring regularly (Figure 1)   

 

When the battery icon indicates the low power level (

), replace both batteries with a brand new 

set of AAA alkaline batteries (1.5V).   

 

Static images may occasionally appear on LCD. Such retained pixel charges do not indicate a faulty 
display, consume the battery power, or affect the instrument

s performance in any way. 

 

Check the expiration dates on batteries before purchase.   

 

Zero-set the instrument after the batteries are replaced. 

 

1. 

Insert a coin in the 

groove on the battery 
compartment cover. Turn 
the coin counterclockwise 
to remove the cover. 

 

 

2. 

Insert batteries, 

observing the correct 
polarity. 

 

 

3. 

Align the cover and 

push it down. 

 
 

 

 

4. 

Close the battery 

compartment cover by 
pushing the cover in with a 
coin in the groove and 
turning it clockwise until it 
stops. 

*Turning excessively may cause 

malfunction. 

 
 
 
 
 

 

Groove 

Approx. 
2mm 

Approx. 0.3mL of sample 

Approx. 0.3mL of water 

Approx. 
2mm 

Figure1 

O-ring 

Mild 

soap

 

Отзывы: