- 16 -
7. Inspection
Regular inspection is recommended to ensure accurate results. Keep a regular schedule, such as the
beginning of each month. Inspect the instrument as needed - when the instrument has been dropped, or
the measurements seem to be abnormal.
The standard liquid is available from ATAGO.
RE-99110 RePo/Po Standard liquid
<Inspection of the instrument with standard liquid>
①Prepare the instrument for basic operations according to the instruction manual.
②Zero set the instrument with room temperature purified water.
③Measure the Angle of Rotation of the standard liquid (2 or 3 times repeatedly).
④If the measurements are within accuracy, the instrument is calibrated correctly.
If the measurements are not within accuracy, review the procedures to measured correctly.
8. Storage and Maintenance
Store the instrument in a dry
place away from direct sunlight.
Exposure to humidity and heat
may damage the instrument.
Do not use organic solvents
(paint thinner, benzene, gasoline,
etc.) on the plastic body case.
(
r
page 1)
Clean and dry the sample stage
thoroughly, following the
“
Cleaning
”
instructions. (
r
page 8)
Store the unit away from direct sunlight
at a stable temperature with as little
fluctuation as possible.