Zero-setting
A calibration point of refractive index (zero point) is programmed.
Always zero-set the instrument when.
•
The instrument is first turned on.
•
The time lapsed since last measurement is more than an hour.
•
The temperature setting is changed.
1.
Touch "MEASUREMENT" on
the home screen.
2.
Apply purified water on the
prism.
4.
Touch "ZERO.
”
3.
Close the cover plate.
5.
"WAITING" will blink (while
the Peltier thermo module is
in operation).
6.
"READING" will blink.
7.
Zero-setting is complete.
※
Touch "QUIT" to abort the
operation.
※
When the wait time function is
turned on, a timer that counts
down from the specified time
interval is displayed.
※
See page 30 "Selecting Operators
and Registering Sample Names"
for instructions on how to change
the measurement mode, number
of measurements to be taken,
user, and sample name.
※
The screen example above is in
the Brix scale.