background image

32

When  the  orange  color  is  not  bright, 

increase  the  quantity  of  light  (Figure 

9-10) (Refer to page 62).

If the orange color (Figure 9-10) is still 

dark in spite of increase of the quantity 

of light or the boundary line is reddish, 

the light of the illuminating lamp is badly 

applied to the lighting window. In such 

the event, adjust the height and angle 

of the illuminating lamp (light guide) so 

that  the  boundary  line  is  more  clearly 

seen.

Looking through the eyepiece, turn the 

measurement  knob  so  that  the 

boundary  line  meets  the  intersection 

point of the cross hairs (Figure 9-11).

Press the SELECT key on the display 

panel of the refractometer.

Such the indications as shown in Figure 

9-12 appear on the display.

Press the 

∨∧ 

key so that the display 

shows  [4]  as  the  selected  function 

mode number. (Figure 9-13)

Cross hairs

Boundary

line

Boundary

line

Orange color

Figure 

9-11

Figure 

9-10

ATAGO

ABBE REFRACTOMETER

SELECT

SET

PRINT

Figure 

9-12

ATAGO

ABBE REFRACTOMETER

SELECT

SET

PRINT

Figure 

9-13

(6) Adjusting reference point

Figure 

9-14

ATAGO

ABBE REFRACTOMETER

SELECT

SET

PRINT

Press the SET key on the display panel 

of the refractometer, and the indication 

of  the  refractive  index  starts  blinking. 

(Figure 9-14)

Again  look  through  the  eyepiece  and 

turn the measurement knob so that the 

boundary  line  meets  the  intersection 

point of the cross hairs.

Again press the SET key on the display 

panel  of  the  refractometer,  and  the 

indication  of  the  refractive  index  stops 

blinking and continues to light. Therefore 

the indicated refractive index is fixed.

After  the  indicated  refractive  index  is 

fixed,  the  indication  on  the  display 

changes as shown in Figure 9-15.

That  is  all  for  adjustment  of  the  reference 

point (calibration).

Содержание 1410

Страница 1: ...Multi wavelength Abbe Refractometer Instruction Manual Cat No 1414 Cat No 1410 1410 E21 PRODUCT REGISTRATION WARRANTY CARD https www atago net registration english registration php ...

Страница 2: ......

Страница 3: ......

Страница 4: ......

Страница 5: ......

Страница 6: ......

Страница 7: ......

Страница 8: ... may result in shortcircuiting the unit smoking or fire ۍ Do not use the power cable if damaged or broken If used fire electrical shock or burns may result For a new power cable contact your ATAGO distributor ۍ When disconnecting the power cable from the outlet make sure to do it by holding the plug If the cable is pulled for disconnection it may not only break the cable but also cause fire and el...

Страница 9: ... If the tube is damaged replace it with new one as immediately as possible If the damaged tube is continuously used the operator may scald himself with hot circulating water ۍ WARNING ۍ The temperature of hot water to be supplied from a circulating constant temperature bath to this refractometer is limited to 50䉝 and the flow rate of the circulating constant temperature bath should be 10 L min at ...

Страница 10: ......

Страница 11: ......

Страница 12: ......

Страница 13: ......

Страница 14: ...things have been contained in the packing case Contact liquid monobromonaphthalene 4mL for setting reference point 1 Contact liquid nD1 78 4mL for practicing solid matter measurement 1 Lighting glass for film measurement 1 Interference filter 589nm 1 Instructions this manual 1 Inspection certificate 1 DR M2 main unit 1 Light source unit 1 AC power cable 1 Light stand 1 Light guide 1 Allen wrench f...

Страница 15: ......

Страница 16: ...switch to turn on the refractometer and press it again to turn it off however the refractometer is turned on off when the light source unit is turned on Constant temperature water circulating nozzles㸸 For measuring sample at a constant temperature these nozzles are used to circulate constant temperature water around the prism For connection of the circulating constant temperature bath refer to pag...

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ......

Страница 29: ......

Страница 30: ......

Страница 31: ...purified water In function mode 䣝4䣟 䢪the mode on setting of the reference point䢫 make the boundary line of the purified water meet at the intersection point of the cross hairs and press the SET key The DR M2 is automatically calibrated by the refractive index of the purified water at the measurement 䢪indicated䢫 temperature Prepare the 589nm interference filter Prepare purified water Do not use the...

Страница 32: ......

Страница 33: ...undary line Boundary line Press the LAMP key of the light source unit The LAMP indication on the display of the light source unit becomes ON Figure 9 7 Look through the eyepiece and you see such the lines as shown in Figure 9 8 ձ ղ Turn off the room light so that the surroundings of the refractometer become slightly dark Turn the eyepiece to bring it into focus 䢪Figure 9 9䢫 Figure 9 7 ATAGO LIGHTS...

Страница 34: ...the selected function mode number Figure 9 13 Cross hairs Boundary line Boundary line Orange color Figure 9 11 Figure 9 10 ATAGO ABBE REFRACTOMETER SELECT SET PRINT Figure 9 12 ATAGO ABBE REFRACTOMETER SELECT SET PRINT Figure 9 13 6 Adjusting reference point Figure 9 14 ATAGO ABBE REFRACTOMETER SELECT SET PRINT Press the SET key on the display panel of the refractometer and the indication of the r...

Страница 35: ...ex of distilled water 5 1 3339 1 3339 1 3338 1 3338 1 3337 1 3337 1 3336 1 3336 1 3335 1 3334 1 3334 1 3333 1 3332 1 3332 1 3331 6 7 8 9 10 11 12 13 14 15 16 17 18 19 21 1 3329 1 3328 1 3327 1 3326 1 3325 1 3324 1 3323 1 3322 1 3321 1 3319 1 3318 1 3317 1 3316 1 3314 1 3313 22 23 24 25 26 27 28 29 30 31 32 33 34 35 1 3310 1 3309 1 3308 1 3306 1 3305 1 3303 1 3302 1 3300 1 3298 1 3297 1 3295 1 3294...

Страница 36: ......

Страница 37: ......

Страница 38: ......

Страница 39: ......

Страница 40: ......

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ......

Страница 45: ......

Страница 46: ......

Страница 47: ......

Страница 48: ......

Страница 49: ......

Страница 50: ......

Страница 51: ......

Страница 52: ......

Страница 53: ......

Страница 54: ......

Страница 55: ......

Страница 56: ......

Страница 57: ......

Страница 58: ......

Страница 59: ......

Страница 60: ......

Страница 61: ......

Страница 62: ......

Страница 63: ......

Страница 64: ......

Страница 65: ......

Страница 66: ......

Страница 67: ...ant up to 150Υ A set of 10 zip ties For water bath connection For Abbe number Ȥd For Abbe number Ȥd For Abbe number Ȥd For Abbe number Ȥe For Abbe number Ȥe For Abbe number Ȥe 450nm to 1100nm Specify desired wavelength With mounting adapter RE 1600 RE 16500 RE 1640 RE 12100 RE 8100 RE 1195 RE 1197 RE 6191 RE 6193 RE 1498 RE 1499 RE 1196 RE 1199 RE 99080 RE 1146 RE 8507 RE 3520 RE 3521 RE 3522 RE 3...

Страница 68: ......

Страница 69: ...f the main unit When it becomes pale green replace the desiccant according to the following procedure TUnscrew the desiccant case from the main unit Figure 18 2 Turn the cap of the desiccant case counterclockwise to remove and empty all particles of the desiccant from the case Figure 18 3 Insert new desiccant particles in the desiccant case Cap the desiccant case tightly and set it in the main uni...

Страница 70: ......

Страница 71: ...2 䢪with the attached test piece at 500 to 650nm䢫 Ȥd Ȥe indication output 䢪with optical interference filter䢫 For printer DP 63䢪䣄䢫䢢䢪optional䢫 Conforming to Centronics standard Refractometer light source unit light guide light stand From 450nm to 1100nm 䢪with interference filters䢫 䢪For measurement in the wavelength range from 681 to 1100nm the near infrared ray viewer is required 䢫 Component units of...

Страница 72: ......

Страница 73: ...ter If your ATAGO instrument requires servicing please contact ATAGO at the following e mail address Europe North America U S A Mexico Central America Costa Rica El Salvador Guatemala Panama Middle East Africa Iran Turkey Saudi Arabia Israel Lebanon South Africa Egypt Asia Oceania South America Argentina Bolivia Brazil Colombia Chile Ecuador Paraguay Peru Uruguay Australia China India Thailand Kor...

Страница 74: ... 66 21948727 9 66 21171549 customerservice atago thailand com TEL 55 16 3913 8400 customerservice atago brasil com TEL 39 02 36557267 customerservice atago italia com 14432 SE Eastgate Way Suite 450 Bellevue WA 98007 U S A TEL 1 425 637 2107 FAX 1 425 637 2110 customerservice atago usa com TEL 86 20 38108256 info atago china com TEL 7 812 777 96 96 info atago russia com TEL 234 707 558 1552 atagon...

Отзывы: