Disassembly of Evolve™ Process Columns
Adjuster Assembly
1.
Detach any pipework from the top sanitary clamp adapter.
2.
Remove flange fixing bolts from the underside of the top tube flange.
3.
Check that adjuster seal is released by turning the seal actuation nut counterclockwise.
4.
Lift adjuster unit out of acrylic tube unit and place it on its’ side on a clean workspace.
CAUTION - Adjuster units are heavy, adjuster unit will roll
5.
Remove sanitary clamp adapter and retain flow tube seal.
6.
Remove the seal adjust nut and spacer and slide the flow tube and seal adjust tube out of the
threaded rod.
7.
Remove the flow cell locking clip and retain. Slide the stainless-steel tube off the flow tube.
8.
Locate the “flats” on the flow tube and using the correct sized spanner remove the flow tube to
expose the flow tube seal.
9.
The adjuster mesh can be removed from the adjuster cell by carefully prizing the bed support
seal from its location groove in the side of the adjuster cell - a small extraction tool is provided.
Note: care must be taken not to damage to seal or the flow cell.
Figure 9 Adjuster Assembly for EvolveTM 100 Process Column