pH Probe Calibration
Wash
Keep probe into Buffer solution
Calibration
Press Cal Key 3 Seconds
5
7pH___Press_CAL
Calibration During 1 minutes
Wait________60s
6
7pH_Quality_100%
Quality Probe
7
Wash
Keep probe into Buffer solution
4pH___Press_CAL
Calibration During 1 minutes
Wait________60s
10
4pH_Quality_100%
Quality Probe
11
Wash
Press Enter Key to save and exit
14
Normal Status
15
Note:
If you have setting Calibrabration = 7 pH the function has 1 point calibrate only 7 pH buffer solution.