Start up
LDM41E and LDM42E Manual
Page 20
ASTECH GmbH
7
Start up
Make sure that all cable ends are protected against short circuit effects before
you turn on power supply!
Connect cable terminals as required for the particular operating mode. To
prevent short circuits, you should seal unused cable ends!
As part of preparative actions, the LDM41E and LDM42E must be properly
installed in the designated working site, oriented onto the target and kept in a
stable position. The target to be measured should preferentially have a
homogeneous, bright surface.
Caution: Do not use any retro reflectors!
The alignment of the LDM41E and LDM42E is facilitated by a Laser beam
13
that
is visible and can easily be turned on at the PC (e.g. command DT starting the
measurement and switch the Laser on, see 8.2).
7.1
Start up with RS232
For starting up, a PC with RS232 data interface or a USB converter for RS232
and a terminal program are required. We recommend the Windows program
LDMTOOL Version 4.0 or higher.
13
depending on ambient light and target conditions
Содержание LDM41E
Страница 1: ...Manual LDM41E and LDM42E Version 1 6 ...
Страница 48: ...Notice LDM41E and LDM42E Manual Page 48 ASTECH GmbH ...