WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.
CONDUCT A TEST
TEST RESULTS
Connect the Interlock Disable Key (P/N 38075) to the signal input
connector on the rear panel of the instrument. If you’re not utilizing a
DUT enclosure (P/N 39064 or 39656) or other safety device, the Interlock
Disable Key is required in order to run a test.
With the instrument set to the desired test type and your DUT correctly
connected to the instrument, you are now ready to start testing.
If the Continuity function is ON and the resistance of the ground circuit
is less than the Max Resistance setting, the green TEST button will
illuminate.
Push the green TEST button on the front panel. The DUT is tested for
a duration equal to the Ramp and Dwell settings.
PASS:
If the DUT passes the test, you will hear a short audible beep and the
display will indicate the test results.
FAIL:
If a failure occurs, you will hear a long audible alarm and the red flashing
indicator will light up. Press the red RESET button to stop alarm.