TEST CONNECTION
Adapter Box Connections
To increase operator safety, you may elect to use an adapter box for products
terminating in either a two-prong or three-prong line cord.
Plug the black lead from the adapter box (P/N 36544) into the CONT.
CHECK terminal located on the front panel.
Plug the white lead (P/N 4040A-08) from the adapter box into the high
voltage output terminal located on the front panel.
Connect the black ground return lead (P/N 2100A-13) to the front panel
return terminal and connect the other end of the lead to the dead metal
on the chassis of the DUT. Check to ensure a solid connection is made
between the DUT and the return clip.
Plug the line cord of the DUT into the adapter box receptacle.
5
WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.
Note:
Be sure to enable the Ground Continuity check before moving forward (see page 3).
CAUTION
HIGH VOL
TAGE
5KV
AC MAX.
6KVDC MAX.
N
L
POWER
CAUTION
5KVAC MAX.
6KVDC MAX.
H.V.
RETURN
TEST
RESET
CONT.
CHECK
DIELECTRIC WITHSTAND TESTER