SECTION 6 DIAGNOSTICS AND TROUBLESHOOTING
______________________________________________________________________
6-5
6.3
BIS VIEW System Messages and Corrective
Actions
When an alarm is triggered, a message appears in the Message Region of the screen. Possible
messages and the recommended operator actions are listed below. It is a good idea to write
down the exact status message number when it occurs and have that information available if
you should need service. See Section 6.5 “What to do if the BIS VIEW Monitoring System
Requires Service.”
BIS VIEW Messages and Operator Actions
Status Messages:
Causes:
Corrective Actions:
None (screen is blank)
Monitor has not completed
its power ON diagnostics
tests.
1.
Monitor may take up to
three minutes to initialize
trend memory. Do not
disconnect equipment or
press keys during this time.
2.
If several minutes have
passed, restart monitor.
3.
Reset monitor.
4.
Replace monitor.
Connect BISx
1.
BISx disconnected.
2.
Defective BISx cable.
3.
Defective BISx.
4.
Defective monitor.
1.
Connect BISx. Verify all
cable connections.
2.
Inspect/repair cable at
connector end.
3.
Replace the BISx.
4.
Replace monitor.
Re-prep Sensor [0013]
1.
Incorrect sensor
application.
2.
Poor sensor
connections.
3.
Sensor Check fails
4.
Defective PIC.
5.
Defective BISx.
1.
Read Instructions on
sensor package and re-prep
sensor.
2.
Check sensor connections.
3.
Re-prep again or replace
sensor. Verify Sensor
Check passes.
4.
Replace the PIC.
5.
Replace BISx.
Содержание BIS VIEW
Страница 2: ......
Страница 4: ......
Страница 20: ...SECTION 1 SAFETY PRECAUTIONS ______________________________________________________________________ 1 8 ...
Страница 60: ...SECTION 4 QUICK REFERENCE GUIDE ______________________________________________________________________ 4 2 ...
Страница 102: ......
Страница 103: ......