Technical Guideline 

In-Line Purge Pump Filter 

Replacement Instructions 

TG-95-0400 

TG-95-0423 

Rev B | 01/12/21 

Page 4 of 5

 

 

Tech Support Website: ASIsupport.com | Parts Order Website: ASI.parts | Tech Support:  [email protected] or 303.407.6072 

8.

 

To reinstall the tubing assembly, make sure the 

arrow on the bowl/filter assembly is facing 

toward the purge pump. Press each end of the 

tubing into the quick connect fittings on the 

sediment bowl and purge pump until a click is 

heard. Adjust the tubing assembly so there is a 

slight “S” bend in the tubing. (Fig 7) 

 

Fig. 7 

9.

 

Reattach the steel-reinforced vacuum tubing to the brass fitting that was removed in step 3. 

 

WARNINGS/CAUTIONS 

In addition to observing the normal precautions associated with standard dental practices and procedures, 

the following additional precautions should be strictly noted and observed during the set-up, operation, and 

maintenance of this system. 

 

 

 WARNING 

QUALIFIED PERSONNEL ONLY 

The product should only be operated by qualified personnel only. The operator bears responsibility for the correct settings 

and proper use of the system. ASI Dental (ASI) cannot be held liable for any malfunction of this product, or performance 

failure and/or its designed or desired utility, nor can ASI be held liable for injuries to persons or animals, in any case when 

the device is misused or not operated, applied or maintained in strict accordance with user/owner instructions set out in the 

operation manual. In the event of any doubt or question, the user is to contact ASI for clarification or assistance. 

 

Improperly maintained or operated systems or instruments may void the associated warranties. 

 

 

 WARNING 

COMPRESSED AIR 

The compressed air system that operates this unit is under pressure. Compressed air can propel dust or loose particles and 

can cause bodily injury or damage. Always turn the system off and bleed off air pressure before attaching or removing air 

lines or accessories or servicing this unit. All air lines should be periodically inspected and replaced if worn or damaged. 

  

If an outside compressed air supply is used to power this unit, the air supply must be regulated to 80 psi or below. Excessive 

air pressure could cause certain components to rupture. 

 

 

 WARNING 

ELECTRICAL VOLTAGE 

This system is powered by high voltage electricity. Like any other electrically powered device, if it is not used properly, it can 

cause electrical shock. Always plug the power cord into an electrical outlet with adequate fuse protection and proper 

grounding. In the event of a short circuit, grounding reduces the risk of shock by providing an escape wire for the electric 

current. Improper grounding of the unit can result in a risk of electric shock. Always unplug the unit before doing any service 

or repair to the unit. 

Содержание TG-95-0400

Страница 1: ...200933 may easily retrofitted with the In Line Purge Pump Filter by purchasing the purge pump filter PN 95 0400 The replacement screen should be replaced at a minimum quarterly 95 0400 In Line Purge Pump Filter Assembly OR 95 0423 Replacement Screen Filter Qty of 4 CONTENTS Parts Included 1 Tools Required 1 Instructions 1 Warnings Cautions 4 PARTS INCLUDED In Line Purge Pump Filter Assembly 1 OR R...

Страница 2: ...cal cord out of the way of the back door Fig 1 Fig 1 3 To gain access to the purge tubing assembly first remove the steel reinforced vacuum pump tubing by unscrewing the tubing clamp and pulling the tubing from the brass fitting Fig 2 Fig 2 4 Before removing the entire tubing assembly from the unit it may be necessary to manually drain any waste from the suction canister Locate the quick connect f...

Страница 3: ...ain from the suction canister completely Fig 4 If only replacing the filter screen skip Step 6 Fig 4 6 Locate the quick connect fitting where the purge tubing meets the sediment bowl Using a flat head screwdriver press on the quick connect fitting then pull the tubing away from the fitting to completely remove the tubing assembly from the unit Fig 5 Fig 5 7 If replacing the filter screen unscrew t...

Страница 4: ...sons or animals in any case when the device is misused or not operated applied or maintained in strict accordance with user owner instructions set out in the operation manual In the event of any doubt or question the user is to contact ASI for clarification or assistance Improperly maintained or operated systems or instruments may void the associated warranties WARNING COMPRESSED AIR The compresse...

Страница 5: ...ion instruments As part of dental procedures particles of amalgam may be suctioned into the dental suction handpieces and collected within the system and trap filter WARNING INFECTIOUS MATERIALS Infectious disease workplace safety protocols to safeguard against cross contamination of infectious disease should always observed When maintaining the suction system or emptying the contents of the sucti...

Отзывы: