Special Wiring –  See “X” Variation    

 

On Unit Label

 

Variation 

Wire Hookup 

  

 

 

Red 

+ Power 

 

 

XTQ 

Black 

Common 

  

 

 

White 

Output

 

 

Red 

+ Power 

 

 

XTG 

Black 

Common 

  

 

 

Green 

Output

 

K8 Transducers – Electrical Connections

Hirschmann Type

 

PIN 1 

=  + Power 

Red

PIN 2 

=    Common 

Black

PIN 4 

=    Output 

Green

Bendix 4-Pin, 6-Pin

Pin A 

=  + Power 

Red

Pin B 

=    Output 

Green

Pin D 

=    Common 

White

Cable Type F2

               

Red 

=  + Power 

White 

=    Common 

N/A

Green 

=    Output 

Cable Type C1

Red    

=   +  Power

White  

=       Common 

N/A

Green  

=       Output

Ratiometric (mV/V)

Cable Type F2

               

Red 

=  + Power 

 

White 

=  – Power 

 

Green 

=  + Output

Black 

=  – Output 

4 Inch Leads

Red*     =     + Power

White  

=     – Power

Green   =     + Output

Blue 

=     – Output 

*Orange  =     + Power, RoHS compliant  

 

 

   versions

Cable Type F2

Red 

=  + Power 

 

White 

=  – Power 

N/A 

Green 

=  + Output

Black 

=  – Output 

Cable Type C1

Red    

=   +  Power

White  

=   –  Power 

N/A

Green  

=   +  Output

Black  

=   –  Output

 

Bendix 4-Pin

Pin A 

=  +Power 

Red

Pin B 

=  +Output 

Green

Pin C  

=  – Output 

Black

Pin D 

=  – Power 

White

Bendix 6-Pin

Pin A 

=  +Power 

Red

Pin D 

=  – Power 

White

Pin B  

=  +Output 

Green

Pin C 

=  – Output 

Black

Pin E 

=  Shunt Cal.

Pin F 

=  Shunt Cal.

Cable Type F2

Red 

=  + Power               N/A

Black 

=  – Power

Cable Type C1

Red 

=  + Power               N/A 

Black 

=  – Power

Hirschmann Type

 

Pin-1 

=  + Power               Red

Pin-2 

=  – Power               Black

Bendix 4-Pin, 6-Pin

Pin A 

=  + Power               Red

Pin B 

=  – Power               Green

Current Output Units 4-20mA

Recalibration Instructions:

1.  Apply 0% Full Scale Pressure.

2.   Set the output using the Zero adjustment potentiometer.

3.  Apply 100% Full Scale Pressure.

4.  Set the output using the Span adjustment potentiometer.

5.  Repeat steps 1 thru 4 as necessary.

3-Wire Voltage

4-20 mA

4-Wire Ratiometric (mV/V)

Ratiometric (mV/V)

Voltage Output Units 1-5, 1-6 Vdc

K2 Transducers – Electrical Connections

K1 Transducers – Electrical Connections

3.0

.625

HEX

6-PIN BENDIX

0.9 SQ.

-------------------

4-PIN BENDIX

0.8 SQ.

3.0

K1, K2 Transducers

0.5

0.34

0.5

1.1

2.1

.625

HEX

K8 Transducers

NOTE: All dimensions are decimal inches

Dimensions

Wiring Diagrams for All Transducers

K1,  K2,  K8
PRESSURE TRANSDUCER 
INSTRUCTION SHEET

+

+

POWER

SUPPLY

(- Power)

(+ Power)

(- Output)

(+ Output)

METER

TRANSDUCER

RED

WHITE

GREEN

BLACK

+

+

POWER

SUPPLY

(Common)

(+ Power)

(+ Output)

METER

TRANSDUCER

+

+

POWER

SUPPLY

(+)

(–)

METER

TRANSDUCER

POWER

OUTPUT

MATING  

CONNECTOR 

CABLE CODING

 CONNECTOR 

TYPE

MATING  

CONNECTOR 

CABLE CODING

 CONNECTOR 

TYPE

MATING  

CONNECTOR 

CABLE CODING

 CONNECTOR 

TYPE