2  Safety 

2.1  General sources of hazards 

Pressure  gauges are pressurized parts where failure can result in hazardous situations. The  selection  of pressure 

gauge should be made in accordance with the rules set out in EN 837-2 & DIN 16003

2.2  Use in accordance with intended purpose 

The devices are only to be used for the intended purpose as described by the manufacturer. 
The devices are used for direct display of differential pressure. 
The integrated switching elements are magnetic spring contacts or inductive proximity switches with a groove design, 

supplied by isolating switch amplifiers. If the set limit values are exceeded, the output circuits are opened or closed. 
For each use scenario, the corresponding set-up regulations must be respected. The use in explosion risk areas is not 

permitted. 

2.3  Operator’s responsibility 

Safety instructions for proper operation of the device must be respected. They are to be provided by the operator for 

use by the respective personnel for installation, servicing, inspection and operation. Risks from electrical energy and 

from the released energy of the medium, from escaping media and from improper connection of the device must be 

eliminated. The details for this are to be found in the corresponding applicable set of regulations, such as DIN EN, 

UVV (accident prevention regulations) and in sector-specific instances of use (DVWG, Ex-. GL,  etc.) the VDE 

guidelines and the regulations supplied by local utilities companies. 
The device must be taken out of service and secured against inadvertently being restarted, if the presumption is that 

risk-free operation is no longer possible (see Chapter 10: Faults). 

Conversion works or other technical alterations to the device by the customer will violate the approval for 

hazardous area and are not permitted. This also applies to installation of spare parts. Possible conversations 

or alterations may only be carried out by the manufacturer. 

The operational safety of the device is only guaranteed where it is used for its intended purpose. The specification of 

the device must be adapted to the medium used in the plant. The limit values indicated in the technical data must not 

be exceeded. 
The safety information detailed in this Operating Manual, existing national regulations for accident prevention, and the 

operator’s internal regulations regarding working, operations and safety must be respected. 
The operator is responsible for all specified servicing, inspection and installation works being carried out by authorized 

and qualified specialists. 
The operator is responsible that the device is used in the correct ATEX zone. 

2.4  Staff qualifications (target group assessment) 

The device may only be installed and started up by specialist staff who are qualified  for  installation, start-up and 

operation of the product. 

 

Specialist staff is people who are able to assess the work assigned to them on the basis of their specialist training, 

their knowledge and experience and their knowledge of the relevant standards, and can identify possible risks.  
For devices in explosion-protected configuration, these staff must have been trained or instructed in, or be authorized 

for, working on explosion-protected devices in potentially explosive plants.  
 

2.5  Signs/Safety markings 

The pressure gauge  and its surrounding packaging carry markings. These markings  show the article number, 

measurement  range, maximum static pressure, maximum overload  and manufacturer. The pressure gauge  can be 

provided with additional signs and safety markings advising on special conditions: 

  Advice on the filling liquid 

  Advice on calibration 

  Ex (for ATEX configuration) 

  Oil-can deleted (if oxygen is used) 

 

Содержание F5509

Страница 1: ...tenance Instruction Manual Differential pressure gauge model F5509 F6509 for industrial application in the following configuration F5509 F6509 ATEX differential pressure gauge without switching contac...

Страница 2: ...nd F6509 without switching contact 5 3 2 F5509 I and F6509 I with inductive proximity switches SJ2 N 6 4 Technical data 7 5 Labeling on the device 7 5 1 Labeling on the device for explosion risk areas...

Страница 3: ...t general safety information 2 and also the following specific advice regarding the intended purposes 2 2 and through to disposal 11 3 contain important safety information which if not followed may re...

Страница 4: ...er will violate the approval for hazardous area and are not permitted This also applies to installation of spare parts Possible conversations or alterations may only be carried out by the manufacturer...

Страница 5: ...clouds and dust layers the ignition temperature must be specified pursuant to ISO IEC 80079 20 2 For dust clouds the maximum allowable medium temperature inside of the measuring instrument must not ex...

Страница 6: ...e in differential pressure between 10 and 90 of the measuring range and a pulse frequency of 0 06 Hz the temperature increase is 10 K To avoid additional temperature increase the devices should not be...

Страница 7: ...tch operating voltage 230 VAC 4 Technical data The detailed technical information can be found in the documents in the Appendix Chapter 12 5 Labeling on the device The label with the serial number and...

Страница 8: ...nnection Distance between ports is 37 mm please consider this when selecting a 5 way manifold 6 3 3 Vent valve The vent valve for the housing is located on the top side If the nipple is pulled out the...

Страница 9: ...tical mounting pressure ports downward When mounted in other orientation max 10 the pointers zero position needs to be adjusted see 8 4 1 Zero point adjustment Connection to be made by authorized and...

Страница 10: ...n site is possible For this proceed as follows Equalize pressure in both chambers Check if internal pressure was built up in the case due to ambient temperature effect Open valve position B wait for p...

Страница 11: ...erformance and recalibration is carried out at regular intervals depending on the application The precise testing cycles should be adjusted in line with the operating conditions and ambient conditions...

Страница 12: ...ty data sheets 11 2 Removal When servicing the gauge the process pressure line must be depressurized the electrical connections isolated from the mains supply and special attention should be taken to...

Страница 13: ...Page 13 of 27 12 2 Declaration of conformity model F5509 F6509 without contacts...

Страница 14: ...Page 14 of 27...

Страница 15: ...Page 15 of 27 12 3 Declaration of conformity model F5509 and F6509 inductive proximity switches...

Страница 16: ...Page 16 of 27...

Страница 17: ...Page 17 of 27 12 4 EU design type test certification for inductive proximity switches of types SJ2 N...

Страница 18: ...Page 18 of 27...

Страница 19: ...Page 19 of 27...

Страница 20: ...Page 20 of 27...

Страница 21: ...Page 21 of 27...

Страница 22: ...Page 22 of 27...

Страница 23: ...Page 23 of 27...

Страница 24: ...Page 24 of 27...

Страница 25: ...Page 25 of 27...

Страница 26: ...Page 26 of 27...

Страница 27: ...Page 27 of 27...

Отзывы: