background image

12

8. Display flashes “APPLY/ REF/ PSI/ THEN/

PRESS/ ENTER/ TO/ START/ OR/ OTHER/
TO/ ABORT”; apply the pressure indicated
in numeric display to gauge and press

 

|

|

display will show “WAIT” and

count down from 6 seconds then briefly
display GOOD then exit into measure-
ment mode, or

9. COMPOUND RANGE GAUGES ONLY RE-

QUIRE ONE ADDITIONAL CALIBRATION
POINT NEAR VACUUM

10. Display flashes “APPLY/ REF/ PSI/ THEN/

PRESS/ ENTER/ TO/ START/ OR/ OTHER/
TO/ ABORT”; apply near vacuum of

–14.000 psi as indicated in numeric

display to gauge and press 

 

|

|

display will show “WAIT” and count
down from 6 seconds then briefly display
“GOOD” then exit into Measurement Mode.

Note:

Recalibration is allowed only if test

parameters are within ±7%.  If outside this
window, the display will indicate “CAL FAIL /
INPUT PRES TOO LOW (HIGH) / PRESS
ENTER TO RETRY / PRESS OTHER TO ABORT”. 

Note: 

“FIELD” option appears only if gauge

has been successfully field recalibrated.

Graph: 

Provides the user the ability to modify pressure
values  dictating the minimum / maximum indi-
cations over the 20 segment bar graph.

1. In menu mode press 

,

until “GRAPH” appears; press 

 

|

|

.

2. Display will indicate “CGOFF” (custom

graph off) [default], “CG ON” (custom
graph on), or “NEWCG” (new custom
graph) press  

to scroll. 

3. To recall the last custom graph entered

when “CG ON” is displayed press 

 

|

|

. Display will briefly show

“DONE” and return to Measurement
Mode.

4. To reset bar graph to full scale range;

when “CGOFF” is displayed press 

 

|

|

Display will briefly show “DONE”, and

 return to Measurement Mode.

5

To enter a new custom bar graph; when
“NEWCG” is displayed press 

 

|

|

.

DG25 I&M006 10197.qxp_I&M006 10197  2/6/17  12:54 PM  Page 12

Содержание DG25

Страница 1: ...4 USA Tel 203 378 8281 Fax 203 385 0402 www ashcroft com All sales subject to standard terms and conditions of sale 2017 Ashcroft Inc I M006 10197 DG25 02 2017 Installation and Operating Instructions...

Страница 2: ...2...

Страница 3: ...permanent damage to the product 3 WARNING READ BEFORE INSTALLATION 1 GENERAL A failure resulting in injury or damage may be caused by excessive overpressure excessive vibration or pressure pulsation e...

Страница 4: ...4 4 GENERAL INFORMATION Keypad Power On Off Enter Zero Up Arrow Tare ZERO Tare Menu Selection Down Backlight Arrow MENU Key presses are short less than 0 5 seconds or long greater than 0 5 seconds...

Страница 5: ...sure display 20 segment pressure range bar graph each segment equals 5 of range Sleep and Backlight Timer symbols Maximum Minimum Pressure and Tare icons 5 character alpha numeric digit display 4 segm...

Страница 6: ...t version and full scale range the unit will then proceed automatically to the Measure ment Mode Zero Function Press ZERO Upon release the numeric display momentarily disappears and ZERO shows in alph...

Страница 7: ...mPa FtH2O InHg cmHG mmHg customer de fined units CUSTU 3 Press to select units and return to Measurement Mode Note Custom units programming see page 8 7 Tare Function Subtracts current pressure value...

Страница 8: ...he digits shall correspond to the last value The decimal point remains at the location chosen in step 1 Press to change the value of the digit between 0 9 Once the desired value is displayed press the...

Страница 9: ...ds prior to gauge shut down Maximum Minimum Displays maximum minimum pressure values this is initiated upon powering the unit or since the values were cleared 1 When in menu mode press to scroll to Ma...

Страница 10: ...urement Mode Note In PRESS mode factory has set 1 hour timer to save battery Feature Back light indicator will blink 10 sec onds prior to light shut off Update Utilized to select the rate at which the...

Страница 11: ...ore values of factory calibration 4 Pressing when FIELD is dis played will restore values from latest field calibration 5 Pressing Enter key when NEW appears will enter the recalibrate mode 6 When NEW...

Страница 12: ...parameters are within 7 If outside this window the display will indicate CAL FAIL INPUT PRES TOO LOW HIGH PRESS ENTERTORETRY PRESSOTHERTOABORT Note FIELD option appears only if gauge has been success...

Страница 13: ...left 2 digits will now begin to flash Press to scroll to a number 0 10 Press to select that number The number entered represents the percentage of full scale to be used as the high end of the graph N...

Страница 14: ...align the unlock icon with the arrow push cover straight in then turn clockwise until the arrow is in alignment with the Lock icon Note Reinstallation of the back cover may cause the unit to read neg...

Страница 15: ...Burst Pressure Vac 2000psi 800 of Span 3000 5000psi 500 7500 25 000psi 300 Environmental Specifications Temperature Storage Batteries removed 20 C to 80 C 4 F to 176 F Batteries installed 20 C to 60...

Страница 16: ...oft Inc 250 East Main Street Stratford CT 06614 USA Tel 203 378 8281 Fax 203 385 0402 www ashcroft com All sales subject to standard terms and conditions of sale 2017 Ashcroft Inc I M006 10197 DG25 02...

Отзывы: