76
Danger, the PTE-2 may not display the full scale psi when
applied (A legacy PTE-1 module with invalid output at full
scale) Contact Customer Service
•
CH1 Low Span / CH2 Low Span
Danger, the PTE-2 may not display the full scale psi when
applied (A legacy PTE-1 module with invalid output at full
scale) Contact Customer Service
• Loop supply overloaded: see troubleshooting section
• Loop supply reset failed: see troubleshooting section
• FRAM write failed: Contact customer service
• FRAM read failed: Contact customer service
• Datalog corrupt: Contact customer service
•
Datalog error corrected: if warning reoccurs contact customer
service
•
Can’t read clock: There is an issue with the internal real time
clock,
• Current date bad: the real time used may not be accurate.
•
Failed to open file/ Failed to close file: SD card failure, card
may be bad or full
System Warnings
• SYS not-calibrated: The PTE-2 is not calibrated, contact cus-
tomer service
•
Out of internal memory: datalog memory is filled, transfer to
SD Card and clear internal datalog to make room.
•
Out of datalog: maximum number of datalogs reached, trans-
fer to SD Card and clear internal datalog to make room.
•
Out of records: present datalog can take no more records,
start a new datalog.
Operation Warnings
•
CH1 zero out of bounds/ CH2 zero out of bounds: input pres-
sure is too high to be zeroed
•
CH1 not a pressure mod / CH2 not a pressure mod: attempt-
ing to zero or tare a module which is not a pressure module
•
CH1 conversion invalid/ CH2 conversion invalid: module
range is not valid for Units selection
•
No SD card present: Attempting to perform an SD card oper-
ation with non-installed SD card
•
SD card locked: attempting to transfer to an SD card but the
write protect tab is engaged.
•
No module to damp: attempting to damp with no module
present
•
Can’t enter MENU (log): The MENU cannot be entered while
datalogging, press the Esc key to halt a datalog when ready.
System Warnings
• Low battery: main battery is low, replace with four fresh AA
alkaline batteries
•
RTC battery low: Real time clock battery is near end of life.
The battery needs to be replaced. This is not a field service
item. Contact customer service.
• Bad msg sent to display: reset PTE-2, contact customer service
Percent Error Operation messages
• % Err: in & out match: refer to section
• % Err: bad input range: refer to section
• % Err: bad output range:
• % Err: ext input of: press V/mA key to turn external channel on
• % Err: CH1 empty: no module is installed in Channel 1 bay
• % Err: CH2 empty: no module is installed in Channel 2 bay
SECTION 16 b.
TROUBLE SHOOTING CONT.
Содержание ATE-2
Страница 1: ...i I M002 10233 ASHCROFT ATE 2 Operation and Maintenance Manual 0 0 5 4 P S I 0 8 6 i n H 2 0...
Страница 2: ...ii...
Страница 50: ...48 SECTION 15 l FIRMWARE UPDATE CONT...
Страница 51: ...49 SECTION 15 l FIRMWARE UPDATE CONT...
Страница 55: ...53 SECTION 15 l FIRMWARE UPDATE CONT Note The File Transfer can take up to 15 Minutes...