32
-0.11
inH20
0.000
volts
-0.011
PSi
Auto–lo
gg
in
g
SECTION 15 f. iii.
Initiate Auto Datalog
SECTION 15 f. iv.
Initiate Manual Datalog
Set lo
g
interval
Desired
:
00001.0 sec
Actual
:
00001.2 sec
Enter to accept
Esc to cancel
“^ “Beneath a digit indicates which character in the time interval
name is being configured.
Press UP or DOWN ARROW to scroll through numeric values
Use left or RIGHT ARROW to move cursor to next desired
position.
When the desired time interval in seconds is displayed, press
ENTER
The HHC will return to Measurement Mode.
*NOTE*
Note: When programming, the display will show “Desired” interval
and “Actual” interval. Actual interval may not match the desired
value. The minimum sample rate changes are based on the
number of inputs installed in the base unit. The sample rate
increase 0.1 seconds per input (0.1, 0.2 or 0.3 seconds). The
actual interval must be an even multiple of this value. For exam-
ple, if there are two pressure modules and a voltage input active,
the minimum sample rate is 0.3 seconds or even multiples of 0.3
seconds. If the desired value entered is 0.1 seconds the actual
value will remain 0.3 seconds.
Initiate Automatic Datalog
To initiate an automatic Datalog after programming desired chan-
nel tag names and logging interval press DATA STORE key. The
display will show a message saying “Auto-logging”
To disable auto logging press either DATA STORE key or ESC key.
Initiate a Manual Datalog
After programming desired channel tag names
Press MENU key
Use DOWN or UP ARROW to highlight “Datalog…”
Press ENTER
SECTION 15 f. ii.
Program Auto Datalog Time
Cont.