Aseptico AEU-40 Endopex V Скачать руководство пользователя страница 5

6.

3.

DISINFECTION AND MAINTENANCE:

1. Main Unit

Wash display and outside of the enclosure with a soft cloth
using a mild detergent. Rinse off the unit by wiping with a
damp cloth. Never submerge unit or saturate with any disin-
fectant fluids.

2. Electrodes

The electrodes must be sterilized between each patient use.
Sterilize the file clip and lip clips at 132º C (270º F) for 5
minutes. Do not autoclave any other part.

3. Cable Assembly

The Cable Assembly may be wiped or dipped in a disinfec-
tant solution. Do not soak the Cable Assembly.

4. Batteries

Replace batteries with fresh cells when unit indicates low
battery when power light is yellow and blinking. From one to
three 9V batteries may be used together, for an estimated
15 hrs of battery life. 

CAUTIONS:

1. Do not use this unit in conjunction with an electric scalpel or

on patients with pacemakers.

2. Replace the battery when the power light starts blinking yel-

low. Accurate readings cannot be assured with batteries that
are low on power.

3. Always remove the batteries for shipping or long-term storage.

4. Use files and reamers with a plastic (resin) handles.

5. Numbers indicating distances are for reference only. The

dentist’s judgment must be used to interpret indication
results. Clinical experience, including knowledge of root
anatomy, is crucial when interpreting indication results.

6. Always verify good connection of the lip clip to the patient’s 

oral mucosa and all connections to the cable assembly 

DISPLAY

SWITCH

SOUND ON/OFF

POWER

SWITCH

LED display gives clear
indication of progress to
the apical foramen.

Adjustable Apex target -
See below for setup

Illustration 1 - Endopex control panel

Illustration 2 - Lip electrode and EZ Hook

TM 

electrode

OPERATION (Continued):

11.DETECTING POST PERFORATION: Connect the file clip to

a large file and make contact with the post. In the case of
post perforation, the Endopex V will indicate a perforation
illuminating the red L.E.D. at the bottom of the indication col-
umn accompanied by a solid audio tone if sound is active.

Содержание AEU-40 Endopex V

Страница 1: ...nstructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the responsibility of the purchaser Accidental misuse inappropriate installat...

Страница 2: ...short indications a Verify the pulp chamber is clean and dry b The canal maybe filled with blood or electrolytic fluids In this case the meter sometimes jumps when the file tip breaks the surface of...

Страница 3: ...Verify that the lip clip has a good connection to the patient s oral mucosa c Verify the file clip metal is clean free of debris and corro sion clean if not d Verify the pulp chamber is clean and dry...

Страница 4: ...nt proximity silence b Near proximity 1 24th pulse chirp c Impending proximity one light bar preceding the Apex indicating 0 5mm 1 3rd pulse beep d Apex 1 2 pulse beep e Perforation solid tone alarm T...

Страница 5: ...yel low Accurate readings cannot be assured with batteries that are low on power 3 Always remove the batteries for shipping or long term storage 4 Use files and reamers with a plastic resin handles 5...

Страница 6: ...turned on The power indicator will turn green indicating the unit is ready for normal opera tion The self test sequence should take only a few seconds to complete 2 Attach the stainless steel lip pro...

Страница 7: ...turned on The power indicator will turn green indicating the unit is ready for normal opera tion The self test sequence should take only a few seconds to complete 2 Attach the stainless steel lip pro...

Страница 8: ...yel low Accurate readings cannot be assured with batteries that are low on power 3 Always remove the batteries for shipping or long term storage 4 Use files and reamers with a plastic resin handles 5...

Страница 9: ...nt proximity silence b Near proximity 1 24th pulse chirp c Impending proximity one light bar preceding the Apex indicating 0 5mm 1 3rd pulse beep d Apex 1 2 pulse beep e Perforation solid tone alarm T...

Страница 10: ...Verify that the lip clip has a good connection to the patient s oral mucosa c Verify the file clip metal is clean free of debris and corro sion clean if not d Verify the pulp chamber is clean and dry...

Страница 11: ...short indications a Verify the pulp chamber is clean and dry b The canal maybe filled with blood or electrolytic fluids In this case the meter sometimes jumps when the file tip breaks the surface of...

Страница 12: ...nstructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the responsibility of the purchaser Accidental misuse inappropriate installat...

Отзывы: