2.

7.

OPERATION FUNCTIONS:

1.

Power Switch

– Controls power on/off to the unit. When

turned on, self-test will initiate and should illuminate a green
power L.E.D. within a few seconds to indicate a properly
functioning unit.

a. A flashing yellow L.E.D. indicates a low battery requiring

replacement.

b. The unit will automatically turn off if it fails the battery of

power on tests. Low batteries will cause self test failures. 

c. If a subsequent attempt to power on succeeds, the unit

will function properly, but should still be returned for serv-
ice if it continues to automatically turn off during self test
with fresh batteries. Low batteries will cause self test fail-
ures.

2.

Display Switch

– Allows the selection of one of three dis-

play modes. The following modes are indicated by the illu-
mination of a corresponding L.E.D. color below the Display
Switch;

a. Single Bar March (Green) – One bar is illuminated to

indicate proximity to the apical foramen. This mode is
most conservative of battery life.

b. Logarithmic March (Yellow) – Further away from the api-

cal foramen, 3 bars will light, mid range proximities will
produce 2 illuminated bars, near proximities, apex indica-
tion and perforation indications will illuminate only one
bar or light. This mode is also conservative of battery life.

c. Persistent On March (Red) – Each illuminated bar

remains on during progression toward the apical fora-
men. This mode is least conservative of battery life, but
certainly gives the most visible indications.

3.

Sound Switch

– Controls sound indication on/off. The state

of the sound indication is indication by the on/off illumination
of an L.E.D. beneath the Sound Switch. When active the
following audio indications will be provided;

a. Distant proximity – silence.

b. Near proximity – 1/24th pulse chirp.

c. Impending proximity (one light bar preceding the Apex

indicating 0.5mm) – 1/3rd pulse beep.

d. Apex – 1/2 pulse beep.

e. Perforation – solid tone alarm.

TROUBLESHOOTING:

1. Nothing happens on the display after the main switch is

pressed.

a. The button should be held down firmly for two seconds to

turn the unit on.

b. Verify the battery is properly installed.

c. Verify the battery is fresh.

2. The power indicator turns red and the unit turns off after

a couple of minutes.

a. Verify the battery is fresh.

b. The first battery of self-tests has failed, wait 5 seconds,

turn the unit back on. If this does not solve the problem,
contact your local dealer or our office listed on the back
page of this manual

3. The power indicator turns yellow and the unit turns off

after a couple of minutes.

a. The second battery of self-tests has failed, wait 5 sec-

onds, turn the unit back on. If this does not solve the
problem, contact your local dealer or our office listed on
the back page of this manual

4. The power indicator is yellow and blinking.

a. Replace the battery.

5. The display gives unstable indications.

a. The file may be following the canal at a sharp angle.

When this happens, the display sometimes "jumps" to a
perforation indication. If this seems the case, retract the
file just enough so the perforation indication stops. This is
the working length.

CAUTIONS (Continued):

including the connection to the unit. A poor connection can
prevent accurate measurements.

7. Always take an x-ray to confirm the results.

8. Do not clean the main unit with Gluteraldhyde, Ethanol, or

similar chemicals to avoid discoloration and surface corro-
sion of the unit. If the unit contacts any of these chemicals,
wipe off immediately with a damp cloth.

9. Indications vary slightly depending on individual tooth char-

acteristics and canal shape; the dentist should apply clinical
judgment on each case.

Содержание AEU-40 Endopex V

Страница 1: ...nstructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the responsibility of the purchaser Accidental misuse inappropriate installat...

Страница 2: ...short indications a Verify the pulp chamber is clean and dry b The canal maybe filled with blood or electrolytic fluids In this case the meter sometimes jumps when the file tip breaks the surface of...

Страница 3: ...Verify that the lip clip has a good connection to the patient s oral mucosa c Verify the file clip metal is clean free of debris and corro sion clean if not d Verify the pulp chamber is clean and dry...

Страница 4: ...nt proximity silence b Near proximity 1 24th pulse chirp c Impending proximity one light bar preceding the Apex indicating 0 5mm 1 3rd pulse beep d Apex 1 2 pulse beep e Perforation solid tone alarm T...

Страница 5: ...yel low Accurate readings cannot be assured with batteries that are low on power 3 Always remove the batteries for shipping or long term storage 4 Use files and reamers with a plastic resin handles 5...

Страница 6: ...turned on The power indicator will turn green indicating the unit is ready for normal opera tion The self test sequence should take only a few seconds to complete 2 Attach the stainless steel lip pro...

Страница 7: ...turned on The power indicator will turn green indicating the unit is ready for normal opera tion The self test sequence should take only a few seconds to complete 2 Attach the stainless steel lip pro...

Страница 8: ...yel low Accurate readings cannot be assured with batteries that are low on power 3 Always remove the batteries for shipping or long term storage 4 Use files and reamers with a plastic resin handles 5...

Страница 9: ...nt proximity silence b Near proximity 1 24th pulse chirp c Impending proximity one light bar preceding the Apex indicating 0 5mm 1 3rd pulse beep d Apex 1 2 pulse beep e Perforation solid tone alarm T...

Страница 10: ...Verify that the lip clip has a good connection to the patient s oral mucosa c Verify the file clip metal is clean free of debris and corro sion clean if not d Verify the pulp chamber is clean and dry...

Страница 11: ...short indications a Verify the pulp chamber is clean and dry b The canal maybe filled with blood or electrolytic fluids In this case the meter sometimes jumps when the file tip breaks the surface of...

Страница 12: ...nstructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the responsibility of the purchaser Accidental misuse inappropriate installat...

Отзывы: