Page 9
Maintenance and
Servicing Instructions
Part Numbers and Assembly Details
Refer to the included Schematic Drawing Set (PN 421204) for component part numbers and
additional service instructions.
Cleaning and lubrication
The external surfaces of the ADU-40CF’s chassis, spout, basin and accessory tray can be cleaned
with a commercial medical/dental disinfectant. Sink hoses contacting waste water must be disinfected
after each session. To minimize the risk of corrosion, chlorine based cleaners should not be used to
clean the interior of the water storage tank or the sink’s legs. After a time, the lines used in the
plumbing may develop a milky mineral deposit. This scale can be reduced by flushing the lines with
a dilute bleach or ammonium solution. External components will be sufficiently maintained by a
cleaning regiment consistent with standard medical and dental practice. Internal components do not
require lubrication or have special cleaning instructions.
Performance verification and Inspection
No specific inspection schedule needs to be maintained for this unit. Components have been selected
for their maintenance free service life. The drain catch should be cleaned after each use. Verifying
that the following conditions are satisfied during operation will ensure proper function. These
conditions should be verified as the final step of any major repair.
1. The green L.E.D. by the spout lights when the unit is connected to a power source.
2. The pump can be heard running when the ON/OFF footswitch is depressed.
3. The tank fills when the footswitch has been depressed for two minutes and the orange L.E.D.
lights to indicate the heating element is operational.
4. The orange L.E.D. goes out when the peak water temperature of 106 degrees Fahrenheit (41
degrees Centigrade) is reached.
5. The orange L.E.D. relights and heating resumes when the heated water falls below a temperature
of 97 degrees Fahrenheit (36 degrees Centigrade).
6. Once primed the intake lines retain their water level when the pump is not operating.