Aseptico ADU-22 Скачать руководство пользователя страница 2

To prevent injury to people and damage to

property, please heed relevant warnings and

remarks. They are marked as follows:

WARNING: 

Serious injury or death 

may result if ignored.

CAUTION:

Damage to property or the

environment may result if

ignored.

NOTE:

Important additional

information and hints.

TABLE OF CONTENTS:

Introduction  . . . . . . . . . . . . . . . . . . . . . . . . .1
Package Contents  . . . . . . . . . . . . . . . . . . . .1
Setting Up the Unit  . . . . . . . . . . . . . . . . . . .2

- Fig. 1 - ADU-22 Cart System  . . . . . . . . .2

- Fig. 2 - Air Supply Connector  . . . . . . . . .2

- Fig. 3 - AA-21 HVE Vacuum System  . . .3

- Fig. 4 - Water Supply Tank  . . . . . . . . . . .3

Operation Functions  . . . . . . . . . . . . . . . . . .4

- Fig. 5 - TA-90D Air/Water Syringe  . . . . .4

- Fig. 6 - Handpiece Controls  . . . . . . . . . .4

- Fig. 7 - Foot Control  . . . . . . . . . . . . . . . .4

- Fig. 8 - Control Console  . . . . . . . . . . . . .5

- Fig. 9 - Water Tank Removal  . . . . . . . . .5

- Fig. 10 - AA-21 HVE Vacuum System  . .6

System Adjustments  . . . . . . . . . . . . . . . . . .7

- Fig. 11 - Handpiece Pressure Controls  .7

Maintenance  . . . . . . . . . . . . . . . . . . . . . . . .8
Warranty  . . . . . . . . . . . . . . . . . . .Back Cover

SPECIFICATIONS:

Dimensions:

Tray:

9.8” W x 13” L

(25 cm W x 34 cm L)

Chassis Height: 23.0” min., adjustable to

38.0” max.

(60 cm min., adjustable to

97 cm max.)

Weight:

15 lbs (7 kg)

Air Supply:

60 - 90 PSI, filtered

P.O. Box 1548, Woodinville, WA 98072

8333 216

th

S

treet S.E. Woodinville, WA 98072

International (425) 487-3157 • Toll Free(800) 426-5913

www.aseptico.com • [email protected]

Содержание ADU-22

Страница 1: ...OPERATION and MAINTENANCE INSTRUCTION MANUAL ADU 22 DENTAL CART SYSTEM...

Страница 2: ...4 Water Supply Tank 3 Operation Functions 4 Fig 5 TA 90D Air Water Syringe 4 Fig 6 Handpiece Controls 4 Fig 7 Foot Control 4 Fig 8 Control Console 5 Fig 9 Water Tank Removal 5 Fig 10 AA 21 HVE Vacuum...

Страница 3: ...the need for a central vacuum pump Congratulations This Aseptico ADU 22 system is engineered to provide many years of reliable service Please read the instructions provided in this manual to receive t...

Страница 4: ...IGH VOLUME VACUUM SYSTEM THREE WAY SYRINGE NWS 8 HDPE WATER SUPPLY TANK HIGHSPEED HANDPIECE CONTROL REMOVABLE INSTRUMENT TRAY 3 The ADU 22 unit must be connected to a dry filtered 60 90 psi air source...

Страница 5: ...stall the HVE vacuum bottle press on locking ring tab Fig 3a push bottle into quick coupler and release tab 5b Hang the vacuum head in the holder at the right hand side of the chassis Fig 3b 6 To inst...

Страница 6: ...ion and the left button for water operation Depressing both buttons will create a mist The syringe features quick change autoclavable tips To remove a tip press on the locking collar surrounding the t...

Страница 7: ...emoving the tank for refilling it must be depressurized by turning the tank 1 4 turn counterclockwise The tank is removed by unscrewing it from its mount at the bottom right rear of the chassis After...

Страница 8: ...e Fig 10 To operate remove the vacuum head from holder and insert a standard oral evacuator tip into end of head extension Depress the on off switch on the side of the hose end for vacuum Waste from t...

Страница 9: ...ndpiece may be adjusted by turning the flow control screw for that handpiece The control screws are located underneath the console next to the HP1 and HP2 handpiece holders Figure 11 Handpiece Pressur...

Страница 10: ...out 5 seconds after every patient and about 20 seconds at the beginning of each day 3 GENERAL CLEANING The cart exterior is designed to make cleaning and the use of prophylactic barriers as easy as po...

Страница 11: ...9 NOTES...

Страница 12: ...or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Cu...