Aseptico ADU-17A2 Express Air 2 Скачать руководство пользователя страница 5

5.

FUSES LOCATED IN 

VOLTAGE SELECTOR 

MODULE

Carefully reinstall fuses into Fuse Holder and slide 

Holder back into Selector Module (Figure 1e). Apply 

pressure until locking tab snaps back into position, 

producing a distinct clicking sound. 

3.  Mount the clip-on Holder to the extended handle of 

the Case, as shown in Figure 3.

Fig. 3

Clip-On 

Holder

Handle

Saliva

Ejector

3-Way 

Syringe

Handpieces

HVE

4.  Attach a sterile handpiece to the ISO Type B (4-Line) 

handpiece connector and place the handpiece in 

Holder (Fig. 3).

5.  Remove the 3-way air/water Syringe from its travel 

clip inside the Case and place it in the Holder.

6.  Position the Waste Tank in the Case with the large 

white connection to the rear left of the instrument 

as shown in Figure 4. Attach the hose connector 

coming from the Vacuum Motor to this large white 

connection. Connect the High Volume Vacuum 

(HVE) hose to the large gray connector located 

on the front of the Waste Tank. Connect the Low 

Volume Vacuum (Saliva Ejector) hose to the black 

connector located on the front of the Tank.

Note:

 When attaching the Saliva Ejector hose to the 

top of the Waste Tank, push in and rotate 1/4 turn 

(see Figure 5).  

7.  Secure Waste Tank by running Bungee Cord from 

hooks on bottom panel overtop Tank lid. Verify that 

the Waste Tank and vacuum lines are properly 

installed. Hang the High Vacuum and Saliva Ejector 

instruments on the Clip-On Holder.

8.  Fill the Water Supply Bottle with clean water and 

connect to the Water Reservoir Connector.

9. 

Attach the foot pedal and place on the floor.

Voltage Selector Module

Fig. 2a

Fig. 2b

Fig. 2c

Fig. 2d

Fig. 2e

120V Indicator

240V Indicator

Содержание ADU-17A2 Express Air 2

Страница 1: ...ADU 17A2 Express Air 2 Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...t if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60 IE...

Страница 3: ...first and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments ma...

Страница 4: ...he Voltage Selector Insert must be installed in the 120 position for 120V 60Hz operation or in the 240 position for 230V 50Hz operation WARNING ALWAYS DISCONNECT POWER TO THE SYSTEM BEFORE OPENING VOL...

Страница 5: ...strument as shown in Figure 4 Attach the hose connector coming from the Vacuum Motor to this large white connection Connect the High Volume Vacuum HVE hose to the large gray connector located on the f...

Страница 6: ...ses air Pressing both buttons simultaneously dispenses an air water mist 5 WATER SUPPLY BOTTLE The Express Air 2 incorporates a self contained pressurized water system This system consists of a 1 lite...

Страница 7: ...d you have made yourself familiar with the system functions there are two operating modes to select from On Off Switch Operation Switch ON activates the vacuum and 3 way air water syringe Foot Pedal O...

Страница 8: ...and between each patient use WATER LINES Disinfect the water lines weekly Prepare a 1 10 bleach solution 1 part household bleach to 9 parts water Remove water reservoir and discard residual water Rep...

Страница 9: ...Check system s power connection Turn On Off switch ON for operation Check that voltage selector module is set for source voltage Check fuse If blown replace with 6A 125V medium time delay fuse for 12...

Страница 10: ...Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM 1 5 Hg 39 6 liters min Water Reservoir Capacity 33 9 fl oz 1 0...

Страница 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...

Страница 12: ...12 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420967 Rev G ECO 15498 04 2022...

Отзывы: