ADU-17A2 Express Air 2 Portable Dental System

OPERATION and MAINTENANCE 

INSTRUCTION MANUAL

Содержание ADU-17A2 Express Air 2

Страница 1: ...ADU 17A2 Express Air 2 Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...t if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60 IE...

Страница 3: ...first and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments ma...

Страница 4: ...he Voltage Selector Insert must be installed in the 120 position for 120V 60Hz operation or in the 240 position for 230V 50Hz operation WARNING ALWAYS DISCONNECT POWER TO THE SYSTEM BEFORE OPENING VOL...

Страница 5: ...strument as shown in Figure 4 Attach the hose connector coming from the Vacuum Motor to this large white connection Connect the High Volume Vacuum HVE hose to the large gray connector located on the f...

Страница 6: ...ses air Pressing both buttons simultaneously dispenses an air water mist 5 WATER SUPPLY BOTTLE The Express Air 2 incorporates a self contained pressurized water system This system consists of a 1 lite...

Страница 7: ...d you have made yourself familiar with the system functions there are two operating modes to select from On Off Switch Operation Switch ON activates the vacuum and 3 way air water syringe Foot Pedal O...

Страница 8: ...and between each patient use WATER LINES Disinfect the water lines weekly Prepare a 1 10 bleach solution 1 part household bleach to 9 parts water Remove water reservoir and discard residual water Rep...

Страница 9: ...Check system s power connection Turn On Off switch ON for operation Check that voltage selector module is set for source voltage Check fuse If blown replace with 6A 125V medium time delay fuse for 12...

Страница 10: ...Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM 1 5 Hg 39 6 liters min Water Reservoir Capacity 33 9 fl oz 1 0...

Страница 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...

Страница 12: ...12 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420967 Rev G ECO 15498 04 2022...