Water Thermometer Calibration
If water temperature is diff erent from temperature shown by ASIN AQUA Salt,
calibrate the thermometer in the water thermometer calibration menu.
Air Thermometer Calibration
If air temperature is diff erent from temperature shown by ASIN AQUA Salt,
calibrate the thermometer in the air thermometer calibration menu.
Testing the pH probe
If the probe meet following qualifi cations, it can be used in the system and it is
functional:
Probe has no visible mechanical damage.
Measured pH value is in tol/- 1,0 (example - water pH is 7,2 and probe
is measuring 7,9 - the tolerance is 0,7 so lower than 1,0 - the probe is OK)
Probe response to positive or negative changes in water or buff er. (example - if
you dive the probe with dry and clean tip to 7,0 pH buff er the 1 minute response
must be at least 90%)
Testing the REDOX probe
If the probe meet following qualifi cations, it can be used in the system and it is
functional:
Probe has no visible mechanical damage.
The redox probe naturally ages so its sensitivity but it should never exceed the
limit tolerance -12% At the buff er test 475 mV it should not measure less than
430 mV.
Probe response to positive or negative changes in the water free chlorine
concentration.
There is no manufacturer of pH and REDOX probes that cover its
products with warranty. ASEKO has decided to cover supplied probes to
its clients by two year warranty period that cover free repair of supplied
probe that will have higher tolerance than above described.
pH - Buff er 7,00 #12065
Redox Buff er 475 mV #12063