background image

Ascon Tecnologic - TRH52 - OPERATING INSTRUCTIONS - PAG. 3

5.  PROBLEMS AND MAINTENANCE

5.1 

Measurements errors

Reading errors can occur if settling time is too short, or be 
caused by steam, sprayed water, air drafts, direct exposure 
to sunlight, or presence of condensation on the sensor.
To obtain accurate measurements the sensor should be left 
to settle in the existing atmosphere for some time.
To reduce the possibility of condensation on the humid-
ity sensor place the probe in ventilated position and where 
there is not a quick and elevation variation of temperature 
(from cold to warm).

5.2 

Cleaning

It is recommended to clean the box only with a cloth welted 
with water or with a detergent neither abrasive nor contain-
ing solvents.
Should be necessary to remove the protective cap, do not 
cause any mechanical stress to the sensors and absolutely 
avoid touching the Humidity sensor surface.

5.3 

Disposal

The appliance (or the product) must be 
disposed of separately in compliance 
with the local standards in force on 
waste disposal.

6.  WARRANTY

The instrument is under warranty against manufacturing 
flaws or faulty material, that are found within 18 months from 
delivery date.  The warranty is limited to repairs or to the 
replacement of the instrument.
The eventual opening of the housing, the violation of the in-
strument or the improper use and installation of the product 
will bring about the immediate withdrawal of the warranty 
effects.
In the event of a faulty instrument, either within the period of 
warranty, or further to its expiry, please contact our sales de-
partment to obtain authorisation for sending the instrument 
to our company.
The faulty product must be shipped to Ascon Tecnologic with 
a detailed description of the faults found, without any fees or 
charge for Ascon Tecnologic, except in the event of alterna-
tive agreements.

7.  TECHNICAL DATA

7.1 

Electrical characteristics

Power supply

: 9... 28 Vdc [Use only SELV type CLASS 2 

power supplies (

 15 W)];

Power consumption

: 20 mA max.;

Maximum load resistance for 4... 20 mA output

:  

150 

Ω

 at 9 Vdc;

Software class

: A;

Relative humidity sensor type

: Current loop;

Relative humidity signal output

: 4... 20 mA (0... 100%RH);

Temperature sensor type

: NTC;

Temperature signal output

: 4... 20 mA (-40... +60°C).

7.2 

Mechanical characteristics

Housing

: Plastic;

Dimensions

: 80 x 80 mm, depth 50 mm;

Weight

: 160 g approx.;

Mounting

: Wall mounting (with sensor towards the bottom);

Connections

: Screw terminals with 5 mm pitch;

Connection wires

: 0.32... 2.0 mm (AWG 22…14);

Recommended multipolar cable

: Cable Ø6.0... 12.5 mm  

(for UL 9... 12 mm/0.35... 0.47 in.);

Enclosure rating

: IP 65;

Action type

: 1;

Environmental rating

: Internal usage, Type 1 enclosure;

Pollution degree

: 2;

Over-voltage category

: I;

Rated impulse voltage

: 330 V;

Period of electrical stress on the insulating parts

: Long 

period (IEC EN 60730);

Shipment and storage temperature

: -40... +80°C.

Storage humidity

: 0... 100 RH% (without icing/condensation).