8.2 General advice for cleaning and disinfecting the handpieces
Important information
You are prohibited from cleaning a handpiece base body under running water. This may
destroy the internal optics and electronics.
The handpieces cannot be sterilized.
Generally, neutral cleaning agents (e.g. Edisonite Super) are to be used!
(The parts are damaged by strongly alkaline solutions!)
Special care is required when cleaning and disinfecting long and narrow cannulas and
blind holes, if existing.
Please observe the special recommendations of the following section, too.
Frequent cleaning and disinfection have low influence on the accessories. The color may
become lighter or changes but this does not affect the use.
The end of the product service life is normally determined by wear, tear and damage
caused by its use.
If the handpieces are not in use for a longer period they must be protected against
contamination.
INSTRUCTIONS
Workplace
The workplace must be clean and disinfected.
Preparations for cleaning
The cleaning of the accessories should be performed as soon as
possible after their use. For this purpose they are to be removed
from the device.
Disassemble accessory units, if required, by following the
instructions provided in the next chapter.
Cleaning:
manual
Equipment: neutral cleaning agent (e.g. Edisonite Super); damp,
lint-free cloth/swabs; soft brush; Q-tip; microfiber wipe
delivered by Asclepion or similar lens cleaning tissues; alcohol or
isopropyl alcohol
1. Procedure for the body of the handpiece:
The accessible surfaces of the handpiece body are cleaned
by using a soft, slightly damp cloth. The cloth must not be
wet or even dripping.
You can use neutral cleaning agent for stubborn dirt.
When doing this make sure that possibly existing optical
surfaces in the body interior are left out in this process!
2. Procedure for optical surfaces:
The accessible optical surfaces must be absolutely clean and
lint-free. Clean them only, if this is not the case.
For doing this, first try to remove the dirt by blowing it away
or by using a brush. If required, the surface is breathed on
and then cleaned with a microfiber (as indicated above). The
Q-tip and the microfiber wipe above the Q-tip can be used
for areas that are difficult to access. Moisten the wipe with a
little bit of alcohol and repeat the process for stubborn dirt.
Содержание MeDioStar
Страница 1: ......
Страница 2: ......
Страница 3: ...User Manual MeDioStar MeDioStar NeXT...
Страница 4: ...Contents 1590 99 002_1 21 Februar 2011...
Страница 6: ...Contents 1590 99 002_1 21 Februar 2011...
Страница 10: ......
Страница 67: ...Application 1590 99 002_1 21 February 2011 57...
Страница 88: ...Disposal 78 1590 99 002_1 21 February 2011 13 EC declaration of conformity...
Страница 94: ......