Operation
40
1590 99 002_1 - 21. February 2011
5.8.3 Activating the laser ready mode and releasing the emission of the laser beam
See also chapter 6
!
After the activation of the Quick start key or of an indication from the indication list, the laser is in
the
mode. Standby is highlighted in white on the
key.
Verify again that
the handpiece has been connected correctly,
all laser safety precautions have been taken.
the door interlock contact functions properly, if you have installed it: Open the
door; it must not be possible to switch the laser to
(an error message is
displayed and must be confirmed).
Set the desired parameters according to the explanations given in the chapters before.
WARNING
The direct and scattered laser radiation can cause serious eye injuries.
During the active laser operation all persons present in the laser room have to wear
protective eye goggles or patient eye shields appropriate for this laser as specified in
chapter 3
Other types of laser protective eye-wear do possibly not provide the
degree of protection needed!
Before pressing the READY key the attending user must check this.
In the
mode you can also activate a laser pulse by pressing the foot switch
unintentionally! Therefore, the laser device must be switched to
in the
treatment pauses and at the end of the treatment.
The laser beam may only be released if the handpiece is precisely directed onto the skin
area to be treated.
In case of dangerous malfunctions the Laser emergency STOP must be activated
immediately.
By pressing the READY/STANDBY key you change to the
mode; the yellow laser warning
label appears on the bottom right of the display. First, the system performs a safety test of the
hardware that takes about 2 sec (see Figure Hardware start).
Figure 35: READY mode of the Laser
Содержание MeDioStar
Страница 1: ......
Страница 2: ......
Страница 3: ...User Manual MeDioStar MeDioStar NeXT...
Страница 4: ...Contents 1590 99 002_1 21 Februar 2011...
Страница 6: ...Contents 1590 99 002_1 21 Februar 2011...
Страница 10: ......
Страница 67: ...Application 1590 99 002_1 21 February 2011 57...
Страница 88: ...Disposal 78 1590 99 002_1 21 February 2011 13 EC declaration of conformity...
Страница 94: ......