Technical Information
Mean,
mmol/L
Pooled
Standard
Deviation
95% Confidence
Interval of Standard
Deviation, mmol/L
Coefficient of
Variation, %
2.6
0.04
0.04 – 0.05
1.7
4.7
0.06
0.06 – 0.07
1.3
7.7
0.12
0.11 – 0.13
1.5
11.2
0.15
0.13 – 0.16
1.3
18.1
0.28
0.26 – 0.30
1.5
Intermediate precision (which includes variability
across multiple days) was evaluated using 3 control
solutions. With each of 3 lots of C
ontour
n
ext
test strips, each control was tested once on each
of 10 instruments on 10 separate days for a total of
300 readings.
Table 5 – System intermediate precision results
for C
ontour
n
ext
EZ meter using C
ontour
n
ext
test strips
Control
Level
Mean,
mmol/L
Standard
Deviation,
mmol/L
95% Confidence
Interval of Standard
Deviation, mmol/L
Coefficient
of Variation,
%
Low
2.37
0.03
0.031 – 0.036
1.4
Normal 7.12
0.09
0.080 – 0.095
1.2
High
20.89
0.36
0.331 – 0.392
1.7
Technical Information
Technical Information
Technical Information
Acceptance criteria in ISO 15197: 2013 are
that 95% of all differences in glucose values
(i.e., between reference method and meter) should
be within ±0.83 mmol/L for glucose values less than
5.55 mmol/L, and within ±15% for glucose values
greater than or equal to 5.55 mmol/L.
Table 3 – System accuracy results for glucose
concentrations between 1.0 mmol/L and
25.6 mmol/L
Within ± 0.83 mmol/L or ±15%
600 of 600 (100%)
A repeatability study was conducted with the
C
ontour
®
n
ext
EZ blood glucose monitoring system
using 5 venous whole blood specimens with glucose
levels from 2.6 to 18.8 mmol/L. Multiple replicates
(n=300) were tested using multiple C
ontour
n
ext
EZ
blood glucose meters and 3 lots of C
ontour
®
n
ext
blood glucose test strips. The following precision
results were obtained.
Table 4 – System repeatability results for
C
ontour
®
n
ext
eZ meter using C
ontour
®
n
ext
test strips
User Accuracy
Precision
A study evaluating glucose values from fingertip
capillary blood samples obtained by 115 lay
persons showed the following results: 100% within
± 0.83 mmol/L of the medical laboratory values at
glucose concentrations below 5.55 mmol/L and
98.8% within ± 15% of the medical laboratory glucose
concentrations at or above 5.55 mmol/L.
Содержание Contour next EZ
Страница 1: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 11: ......
Страница 13: ......
Страница 14: ......
Страница 15: ......
Страница 16: ......
Страница 17: ......
Страница 18: ......
Страница 19: ......
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 23: ......
Страница 26: ......
Страница 31: ......
Страница 34: ......