CST. CO., LTD.
14
4.2.5 Set the function of Progressive and Oscillation
1) Progressive increasing of ROM
<Fig 4.3>
● Until reached to programmed [Upper(flexion) and Lower(extension) limit] angle, the
function of [Progressive] increase flexion and extension angle progressively by 1° for
each round trip of the range of exercise.
● If touch the [Progressive, Oscillation] display area(6) at the screen <Fig 4.1>, the
screen is changed to <Fig 4.2>, [Progressive] is selected. If touched twice, [Oscillation]
will be selected.
● Touch the [Angle] or [Counter] display area, and change the angle or the number of
the [Progressive] with [Up/Down]arrow(4).
● [Angle] is the range of angle to be done the [Progressive] function at the Upper and
Lower limit. [Counter] is the number of repetitive movement every 1° increment.
● The range between the [Upper and Lower limit] angle should be 3 times larger than
the programmed [Progressive] angle.
● When the [Progressive] function is used, the [Upper and Lower limit] angle is adjusted
automatically if the programmed range between [Upper and Lower limit] angle is not 3
times larger than the programmed [Progressive] angle.
● With the [Progressive] angle is already programmed, if the range between [Upper and
Lower limit] angle is set to less than 3 times the [Progressive] angle, then the
[Progressive] function is automatically disabled.
● Setting example
● [Progressive] is disabled after completion of function.
< Progressive Condition >
Upper limit angle : 140°
Lower limit angle : 0°
Progressive angle(PA) : 15°
Progressive Counter : 3